Single Slide Automated Gram Stainer User Manual
16
Art specifications
:
IFU/Manual - 8"Wx10"H
Do not alter art or size.
All art prints PMS 185 Red and
100% Black.
AMS1800062
Rev. 0 2018-06
9.3.2 Press [ 2 ]
on the key pad to select
Clean Cuvette
.
9.3.3 Press [ 2 ]
on the key pad to select
Scrub Stain Lines
9.3.4 Press [ 1 ]
on the key pad to select
Purge
, this will empty the lines of stains.
9.3.5
Move the Crystal Violet, Gram’s Iodine, and Safranin cannulas into one bottle of System Cleaner (Cat. No. CHB-AGS-SC), or a bottle containing at least
500 mL of Decolorizer. The cleaner is used to clean the internal portions of the reagent lines, ports, cuvette, and photo optics of the unit.
9.3.6 Press [ 1 ]
on the key pad to select
Scrub
The unit will automatically prime the system cleaner into the reagent lines and then begin the scrub sequence.
9.3.7
During the scrub sequence use the
Cuvette Swabs
(Cat. No. CHB-GS-CS) to manually scrub any residue build up, off the side of the cuvette chamber.
NOTE:
When swabbing the cuvette, be cautious not to cause an overflow, as that can affect the integrity of the machine.
9.3.8
Once the scrub cycle has completed, wipe off the reagent cannulas with an alcohol pad and return the three stain lines to the proper stain bottles.
9.3.9 Press any key
on the key pad to prime the reagent lines with reagents. The unit will make an audible alert when the Scrub Cycle is
complete and ready to resume use.
NOTE:
To return to the main menu, Press [ * ] on the key pad.