9.0 MAINTENANCE
9.1 Wash Cycle
The Wash feature for the Single Slide Automated Gram Stainer is required every 20 slides or every 8 days.
The unit runs water and decolorizer through the cuvette and with agitation cleans the internal surface of the cuvette and tubing.
9.1.1 Press [ 1 ]
on the key pad to select
Run
9.1.2 Press [ 2 ]
on the key pad to select
Clean Cuvette
9.1.3 Press [ 1 ]
on the key pad to select
Wash Cuvette
9.1.4 Press any key
on the key pad to run the Wash Cycle
When complete, there will be an audible alert notifying the technician that the machine is ready to resume staining.
9.2 Bleach Cycle
The Bleach feature for the Single Slide Automated Gram Stainer is required every month or 600 cycles, whichever occurs first. The unit runs a 10% bleach solution
through the reagent and water lines to clean the internal surface of the tubing and to prevent debris build up in the waste line.
9.2.1 Press [ 1 ]
on the key pad to select
Run
9.2.2 Press [ 4 ]
on the key pad to select
Purge
9.2.3
Remove the reagent lines (Crystal Violet, Gram’s Iodine, and Safranin) from their stain containers, and
wipe off
the residual reagents
from the ends of the cannulas or rinse them off in water.
9.2.4
In the same water container that supplies the instrument with water,
make a 10% bleach solution
.
9.2.5 Submerge
all three of the reagent lines and the D.I. Water line into the DI water container that is full of a
10% bleach solution
.
9.2.6 Press [ * ]
on the key pad to return to the Main Menu
9.2.7 Press [ 1 ]
on the key pad to select
Run
9.2.8 Press [ 3 ]
on the key pad to select
Prime
9.2.9 Press [ * ]
on the key pad to return to the Main Menu
9.2.10 Press [1]
on the key pad to select
Run
9.2.11 Press [1]
on the key pad to select
GRAM STAIN