70-00569-02 G
Powerheart® G5 Automated External Defibrillator
D-4
Electromagnetic Emissions Standards Compliance
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec.
Not applicable
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
80 A/m
Power frequency magnetic fields should
be at levels no higher than those
characteristic of a typical location in typical
heavy industrial and power plants and the
control rooms of H.V. sub-stations.
Note:
U
T
is the a.c. mains voltage prior to application of the test level.
Conducted RF
3 Vrms
Not Applicable
IEC 61000-4-6
150 kHz to 80 MHz outside
ISM bands
a
10 Vrms
150 kHz to 80 MHz in ISM
bands
a
Not Applicable
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment—
guidance
Summary of Contents for Powerheart G5
Page 1: ...User Guide 70 00569 02 G Cardiac Science Powerheart G5 Automated External Defibrillator...
Page 2: ......
Page 3: ...USER GUIDE POWERHEART G5 AUTOMATED EXTERNAL DEFIBRILLATOR 70 00569 02 G...
Page 8: ...Contents 70 00569 02 G vi Contents...
Page 14: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 1 6 About the AED...
Page 22: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 2 8 Steps of a Rescue...
Page 36: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 4 4 AED Features...
Page 93: ......
Page 94: ......
Page 95: ......