Diagnostic mode messages
MAINTENANCE REQUIRED
CONTACT TECH SUPPORT
The AED detects a condition
that has no negative effect on
the ability to perform a rescue.
The AED can be used to
perform a rescue.
Contact Cardiac Science
Technical Support or your
local representative.
TEMP TOO HOT
ADJUST STORAGE TEMP
The AED is hotter than its
permitted storage temperature.
While this condition should be
remedied as soon as possible,
the AED may be used to
perform a rescue.
Move the AED to a cooler
location.
TEMP TOO COLD
ADJUST STORAGE TEMP
The AED is cooler than its
permitted storage temperature.
While this condition should be
remedied as soon as possible,
the AED may be used to
perform a rescue.
Move the AED to a warmer
location.
BATTERY LOW
CHECK BATTERY
The battery charge is low,
although a rescue can continue
for approximately 9 more shocks.
While this condition should be
remedied as soon as possible,
the AED can be used to perform
a rescue.
Replace the battery before
the next rescue.
Text display
Voice prompt
Line 1
Line 2
Situation
Resolution
Summary of Contents for Powerheart G5
Page 1: ...User Guide 70 00569 02 G Cardiac Science Powerheart G5 Automated External Defibrillator...
Page 2: ......
Page 3: ...USER GUIDE POWERHEART G5 AUTOMATED EXTERNAL DEFIBRILLATOR 70 00569 02 G...
Page 8: ...Contents 70 00569 02 G vi Contents...
Page 14: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 1 6 About the AED...
Page 22: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 2 8 Steps of a Rescue...
Page 36: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 4 4 AED Features...
Page 93: ......
Page 94: ......
Page 95: ......