70-00569-02 G
Powerheart® G5 Automated External Defibrillator
5-4
Troubleshooting
Maintenance and service messages
These messages may appear during a periodic self test or during a rescue at any prompt level.
Use the following table to resolve any messages that the AED might display.
Text display
Voice prompt
Line 1
Line 2
Situation
Resolution
Battery low.
BATTERY LOW
The battery charge is low,
although a rescue can continue
for approximately 9 more shocks.
Replace the battery before
the next rescue.
REPLACE BATTERY NOW
Occurs when the lid is opened to
perform a rescue and the battery
charge is low.
The battery charge is too low to
support a rescue. In addition,
the following events occur:
•
The Rescue Ready indicator
turns red
•
The AED beeps once every
30 seconds
Replace the battery before
continuing with the rescue.
If the battery charge is
completely depleted,
all AED activity ends.
Open lid to
continue rescue.
OPEN LID
TO CONTINUE RESCUE
The lid is closed during a rescue.
The prompt repeats for 15 seconds.
Ensure that the AED’s lid is fully
open.
Make sure pad
connector
is plugged
into AED.
CHECK CONNECTOR
IS PLUGGED INTO AED
The defibrillation pads have
become disconnected from
the AED.
Ensure that the pads are
securely plugged in to the
AED. Resume the rescue.
Summary of Contents for Powerheart G5
Page 1: ...User Guide 70 00569 02 G Cardiac Science Powerheart G5 Automated External Defibrillator...
Page 2: ......
Page 3: ...USER GUIDE POWERHEART G5 AUTOMATED EXTERNAL DEFIBRILLATOR 70 00569 02 G...
Page 8: ...Contents 70 00569 02 G vi Contents...
Page 14: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 1 6 About the AED...
Page 22: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 2 8 Steps of a Rescue...
Page 36: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 4 4 AED Features...
Page 93: ......
Page 94: ......
Page 95: ......