70-00569-02 G
Powerheart® G5 Automated External Defibrillator
5-8
Troubleshooting
PADS EXPIRED
REPLACE PADS
The AED detects that the
connected defibrillation
pads are older than their
“use by” date.
CAUTION:
Use of pads that are
damaged or expired may result
in improper AED performance.
Replace the defibrillation
pads.
PADS USED
REPLACE PADS
The AED detects that the
connected defibrillation pads
have already been used in
a rescue.
WARNING!
Used pads may not
adhere properly to the patient.
Improper pad adhesion may
result in skin burns. Improper
pad adhesion may result in
improper AED performance.
Used pads may cause patient-
to patient contamination.
Replace the defibrillation pads.
CHECK PADS
The AED detects an issue with
the defibrillation pads.
Ensure that the connector is
securely plugged into the
AED. Replace the pads if
necessary.
NEXT
The AED detects more than
one error.
Press the lit button to view
the next error.
CLEAR
The AED displays a TEMP TOO
HOT or TEMP TOO COLD error.
Press the lit button to
remove the error message
from the AED.
Text display
Voice prompt
Line 1
Line 2
Situation
Resolution
Summary of Contents for Powerheart G5
Page 1: ...User Guide 70 00569 02 G Cardiac Science Powerheart G5 Automated External Defibrillator...
Page 2: ......
Page 3: ...USER GUIDE POWERHEART G5 AUTOMATED EXTERNAL DEFIBRILLATOR 70 00569 02 G...
Page 8: ...Contents 70 00569 02 G vi Contents...
Page 14: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 1 6 About the AED...
Page 22: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 2 8 Steps of a Rescue...
Page 36: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 4 4 AED Features...
Page 93: ......
Page 94: ......
Page 95: ......