Powerheart® G5 Automated External Defibrillator
70-00569-02 G
A-5
Press pads firmly to patient’s
bare chest.
PRESS PADS FIRMLY
TO CHEST
Prompts when better pad
contact to the patient’s skin
is required.
X
X
X
Table A-3: Pad prompts (continued)
Text display
Prompt level
Voice prompt
Line 1
Line 2
Situation
Adv
Std
Bas
Table A-4: Analysis
Text display
Prompt level
Voice prompt
Line 1
Line 2
Situation
Adv
Std
Bas
Do not touch the patient!
Analysing heart rhythm.
Please wait.
DO NOT TOUCH PATIENT
ANALYSING RHYTHM
Repeats until analysis of the
patient’s cardiac rhythm is
completed. This prompt will
be interrupted if the AED is
ready to shock.
X
X
Do not touch the patient!
Analysing rhythm.
DO NOT TOUCH PATIENT
ANALYSING RHYTHM
Repeats until analysis of the
patient’s cardiac rhythm is
completed. This prompt will
be interrupted if the AED is
ready to shock.
X
Shock advised. Do not touch
the patient.
SHOCK ADVISED
DO NOT TOUCH PATIENT
Notifies the rescuer that
a shockable rhythm has
been detected and the
AED is preparing to deliver
a defibrillation shock (charging).
X
X
X
Shock not advised.
SHOCK NOT ADVISED
Notifies the rescuer when the
AED detects a non-shockable
rhythm.
X
X
X
Summary of Contents for Powerheart G5
Page 1: ...User Guide 70 00569 02 G Cardiac Science Powerheart G5 Automated External Defibrillator...
Page 2: ......
Page 3: ...USER GUIDE POWERHEART G5 AUTOMATED EXTERNAL DEFIBRILLATOR 70 00569 02 G...
Page 8: ...Contents 70 00569 02 G vi Contents...
Page 14: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 1 6 About the AED...
Page 22: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 2 8 Steps of a Rescue...
Page 36: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 4 4 AED Features...
Page 93: ......
Page 94: ......
Page 95: ......