70-00569-02 G
Powerheart® G5 Automated External Defibrillator
B-6
Technical Data
Sensitivity and specificity
of Rhythm Detection
Shockable Rhythm—VF: Meets IEC 60601-2-4 requirement and AHA
recommendation of Sensitivity of >90%
Shockable Rhythm—VT: Meets IEC 60601-2-4 requirement and AHA
recommendation of Sensitivity of >75%
Non-shockable Rhythm—NSR: Meets IEC 60601-2-4 requirement (>95%)
and AHA recommendation (>99%) of Specificity
Non-shockable Rhythm—Asystole: Meets IEC 60601-2-4 requirement and
AHA recommendation of Specificity of >95%
Non-shockable Rhythm—all other rhythms: Meets IEC 60601-2-4
requirement and AHA recommendation of Specificity—all other rhythms
of >95%