Figure 7. Disinfectant Transfer Pump
Empty the reservoir using Transfer Pump and Tube set
Note: Some local, regional, or national agencies require neutralization of the
solution prior to disposal. Use the MEDIVATORS MediNewt™ system if
disinfectant neutralization is required.
1. Plug the Transfer Pump into a standard power outlet.
2. Connect the Transfer Tube set between the pump and the reservoir.
3. Place the rigid end of tube “A” into the drain or solution container and the opposite end to the outlet of
the transfer pump.
• Connect tube “B” between the inlet of the pump and the reservoir.
4. Turn the pump
ON
at the switch.
Note: Ensure that there are no leaks. If leaks are seen, turn the pump OFF.