22
Figure 3. Transfer Pump Connected to Prefilter and Solution
• Insert the inlet tubing of the transfer pump into a one gallon disinfectant container.
• Press
START
on the reprocessor.
• Turn the transfer pump
ON
until the pre-filter canister is filled up and the disinfectant solution starts
flowing continuously in the basin.
• Press the
STOP
button on the reprocessor.
• Turn the transfer pump
OFF.
• Allow the filters to soak for the time recommended for the high level disinfectant being used to achieve
decontamination/sanitization.
• Reconnect the water line to the inlet side of the filtration system.
• Turn the water
ON
.
• Press
START
and allow the wash cycle to run in order to rinse the disinfectant out of the basin. When
all of the water has drained out of the sink, press
STOP
.
Filtration