51
•
MANUAL “AIR” Cycle:
Sets the reprocessing cycle to perform a separate Air purge cycle.
• Press the
MANUAL
“
AIR
” key multiple times to cycle through and select the desired air purge time
of 0, 3, 5 and 10 minutes. The Air LED light will now illuminate.
• Press START/STOP key to begin the air cycle. The
START/STOP
light will illuminate green, the
AIR
key will begin blinking, and the display screen will show AIR and cycle time remaining in
minutes.
• The CER OPTIMA
®
system cover can be opened during the cycle; however, be careful of spraying
fluids. Wearing appropriate PPE is required.
• Upon cycle completion, the audible tone will sound, the
START/STOP
light will turn red, and the
display screen will show Cycle Completed. The printer will automatically generate a printed cycle
report (if auto-print enabled), or press PRINT to manually generate a cycle report.
• Note: The cycle data is retained in the CER OPTIMA system until either the next cycle is started, or
until system power is shut off.
• The completed cycle data will be automatically transferred to the optional CER OPTIMA Data
Management Program (if connected) at the completion of the cycle.
•
MANUAL “ALCOHOL” Cycle
: Sets the reprocessing cycle to perform a separate Alcohol purge cycle.
• Press the
MANUAL
“
ALCOHOL
” key multiple times to cycle through and select the desired air
purge volume of:
•
The CER-1 OPTIMA allows for 0, 30 or 60cc’s volume
•
The CER-2 OPTIMA allows for 0, 30, 60 or 120cc’s volume
• The Alcohol LED light will now illuminate.
• Press
START/STOP
key to begin the alcohol cycle. The
START/STOP
light will illuminate green,
the Alcohol key will begin blinking, and the display screen will show
ALCOHOL
and cycle time
remaining in minutes.
• The CER OPTIMA system cover can be opened during the cycle; however, be careful of spraying
fluids. Wearing appropriate PPE is required.
• Upon cycle completion, the audible tone will sound, the
START/STOP
light will turn red, and the
display screen will show Cycle Completed. The printer will automatically generate a printed cycle
report (if auto-print enabled), or press
to manually generate a cycle report.
• Note: The cycle data is retained in the CER OPTIMA system until either the next cycle is started, or
until system power is shut off.
• The completed cycle data will be automatically transferred to the optional CER OPTIMA Data
Management Program (if connected) at the completion of the cycle.