Without using the Disinfectant Transfer Pump.
• Fill each housing with approximately 40-ounces of high level disinfectant.
• Screw the housings into the bracket. Make sure that the filters are well seated into the peg in the lid of
the housing. If the pre-filter and the gasket are not properly seated, the housing will not screw into the
bracket.
• Allow the filters to soak for the time recommended for the high level disinfectant being used to achieve
decontamination/sanitization.
• Turn water
ON
.
• Press
START
on the reprocessor.
• As soon as HLD begins to flow into the basin, press
STOP
.
• Allow the system to stand filled with disinfectant for the time recommended for the high level
disinfectant being used to achieve decontamination/sanitization.
• Press
START
and allow the wash cycle to run in order to rinse the HLD out of the basin. When all of
the water has drained out of the sink, press
STOP
.
Reprocessor Installation
Air Filter
Insert the air filter into the connection point label “
AIR FILTER
” located on the reprocessor’s back panel.
NOTE: the air filter can be attached directly to the CER OPTIMA
®
unit, or it can be attached using the
Air Filter connection tubing. The air filter connection tubing may allow easier accessibility to the air filter,
depending on the use location of the CER OPTIMA unit. Refer to the “
Maintenance and Troubleshooting
”
section for details on changing the air filter.
Caution!
The CER OPTIMA system is designed to be used with the
MEDIVATORS air filter. Failure to use the air filter could result in unit damage.
Barcode Reader
Mount the barcode reader bracket onto the CER OPTIMA unit right side panel using the two mounting screws.
Insert the barcode reader electrical connection end into the CER OPTIMA unit connection point labeled
“Barcode” located on the reprocessor’s back panel. Place the barcode reader into the holder.
Data
If using the optional CER OPTIMA Data Management Package, then insert the USB data cable into the
connection point labeled “
DATA
” located on the reprocessor’s back panel.
!