62
Air Filter
Changing the Air Filter:
The CER OPTIMA
®
Endoscope Reprocessing System uses an air filter to ensure only filtered air is delivered
into the unit and into the endoscope. The air filter condition is to be monitored weekly and replaced every three
(3) months.
To replace the air filter
Step 1:
Locate a new air filter.
Step 2:
Disconnect the used air filter from the CER OPTIMA system back panel by pressing the quick-release
fitting.
Step 3:
Insert the color-matching connector of the new air filter into the CER OPTIMA system air filter
connector; push in until a snap-lock is heard. Verify the new air filter is firmly secured in place.
Step 4:
Record the date of the air filter change in the Filter Log, and discard the used filter.
Figure 20