72
|
Message
Description
SSG Data Er
Alarm indicating that the SSG Database was out of range. DSD and SSG data do not
match.
SSG Dose Err
Alarm indicating that the SSG is not indicating completion prior to DSD step advancing
(SSG lost database or program step error)
SSG Fill Err
Alarm indicating that the DSD has advanced to the next step prior to SSG fill
completion (program step error)
SSG HW Err
Alarm indicating that the SSG has experienced a hardware error (includes capacitive
sensor missing)
SSG Missing
Alarm indicating the DSD detected a failure in the SSG Communication (DSD doesn’t
see the SSG)
SSG Rsvr Err
Alarm indicating that an SSG dosing reservoir sensor has remained high (not empty)
after disinfect cycle
Basin In Use
Indicates that the selected operation cannot be performed because the station is not
currently idle
Task Unfin
Alarm indicating that the SSG is reporting a Task Not Finished Error; fill or dose did not
complete
Time Err
Time base error, can only be cleared in Diagnostics 70
Water T High
Incoming water temperature exceeding maximum allowed
Water T Low
Incoming water temperature lower than minimum allowed