DISINFECT WATER LINES
Caution!
Ensure the restrictor adapter provided with the installation kit is connected in both
basins before performing this procedure.
Note:
The disinfectant will remain in the lines for the pre-programmed water line disinfection
time. Upon completion of a water line disinfection cycle, purge the on-board .2 micron water
filter housing of air.
Note:
Both stations must be idle to perform this procedure.
1. On the control panel, press
SETUP
.
2. Type
6
on the keypad, then press
ENTER
.
3. Press
START
. “Attach Restrictor” displays on the screen.
4. Attach restrictors in both basins.
5. Press
START
.
6. Upon completion of the water line disinfection cycle, purge
air from the on-board .2 micron water filter housing (for
complete instructions, see
Periodic Maintenance
).
Figure 6.2
- Attach Restrictor
Figure 6.3
- Location of Restrictor