29
CHECK ALCOHOL AND DETERGENT LEVELS
Note:
This procedure is applicable if alcohol or detergent are utilized.
Caution!
For the alcohol and detergent bottles; remove, refill and replace each bottle
(alcohol and detergent) separately to ensure the correct fluid is in the reservoir.
Note:
Allow 1 inch of space at the top of the detergent and alcohol reservoirs to
accommodate the reservoir sensors.
1. Check the alcohol reservoir (if utilized) for proper
alcohol level.
2. If necessary, remove reservoir bottle and add
alcohol to fill line.
3. Re-attach bottle to cap.
Figure 4.2
- Alcohol Reservoir
4. Check the detergent reservoir (if utilized) for proper
detergent level.
5. If necessary, do the following to add detergent:
a. Remove reservoir bottle.
b. Wipe the reservoir cap, bracket and bottle of
any spilled fluids.
c. Rinse the bottle thoroughly with hot water.
d. Add detergent to the fill line.
e. Re-attach bottle to cap.
Figure 4.3
- Detergent Reservoir