11. “Alcohol” displays on the screen. To set the alcohol air purge time, do the following:
a. Type the two-digit number of minutes, then press
ENTER
.
b. Type the two-digit number of seconds, then press
ENTER
.
12. “Alcohol Inject” displays on the screen. To set the alcohol inject time, type the two-digit number of seconds,
then press
ENTER
.
The volume of alcohol is controlled by the number of seconds entered on the screen, up to a maximum of
59 seconds.
1 second injection
= 3 ml alcohol
13. “Air” displays on the screen. To set the air cycle time, do the following:
a. Type the two-digit number of minutes, then press
ENTER
.
b. Type the two-digit number of seconds, then press
ENTER
.
14. The custom program setting is complete. Record the settings on the next page (
Custom Program Reference Chart
).