• ECF Sensor Check (IREDs on)
P
REPARATIONS
:
1) Attach the IRED positioning tool to the eyepiece.
2) On the ECF Adjustment menu, select [F3] ECF Sensor Check (IREDs on).
C
HECK
P
ROCEDURE
:
1) After a few moments, the waveforms
as shown on the right will appear.
Four pillars on the left and right (total
of eight) should be displayed
alternately. The pillars are the IREDs'
light reflecting off the IRED
positioning tool's eye. If you do not
see the four pillars, it means the IRED
is not lit. The ECF sensor is therefore
faulty.
• ECF Sensor Check (Position Check)
P
REPARATIONS
:
1) Securely attach the IRED positioning tool to the eyepiece.
2) Shine a light source (EV12) on the eyepiece.
3) On the ECF Adjustment menu, select [F4] ECF Sensor Check (Position Check).
C
HECK
P
ROCEDURE
:
1) After a few moments, the numbers as
shown on the right will appear. Check
if the numbers are close to the target
values. These numbers indicate the
position of the holes in the IRED
positioning tool. When the ECF unit is
operating normally, the numbers
displayed will be close to the target
values. If the numbers are far off from
the target values 1.5 or more), do the
adjustment again and check again. If
the numbers are the same as before
the adjustment, the ECF sensor is
probably faulty.
Part 4: Electrical Adjustment
4-41
NAC_C4
ECF check
Observe the waveforms with covering eyepiece or
putting light with EF-8000.
If each waveform keeps flat, condition is normal.
ECF Check
Attach the mask holder and set the light source to EV12.
The values below are position of three mask holder hole.
Bracketed values are standards.
I f a c t u a l v a l u e s a r e g r e a t e l y d i f f e r e n c t f r o m
t h e s t a n d a r d s , r e a d j u s t m e n t i s n e c e s s a r y
T o r e t u r n t o E C F m e n u , p r e s s r e t u r n .
Xaxis 000.000 (-40.0) 000.000 (0.0) 000.000 (+40.0)
Yaxis 000.000 (0.0)
000.000 (0.0) 000.000 (0.0)
Interval = 00.000 (-4.94)
Summary of Contents for C12-8453
Page 7: ...Part 1 General Information ...
Page 12: ...Fig 1 2 Three External Views Part 1 General Information 1 5 ...
Page 53: ...Part 1 General Information 1 46 Fig 1 23 Three External Views ...
Page 67: ...Part 2 Technical Information ...
Page 103: ...Part 3 Repair Information ...
Page 104: ...This page intentionally left blank ...
Page 117: ... MEMO Part 3 Repair Information 3 13 ...
Page 123: ... MEMO Part 3 Repair Information 3 19 ...
Page 137: ... MEMO Part 3 Repair Information 3 33 ...
Page 148: ...Part 4 Electrical Adjustment ...
Page 195: ...Part 5 Parts Catalog ...
Page 196: ...This page intentionally left blank ...
Page 198: ......
Page 200: ......
Page 202: ......
Page 204: ......
Page 206: ......
Page 208: ......
Page 210: ......
Page 212: ......
Page 214: ......
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 237: ...Part 6 Electrical Diagrams ...
Page 271: ......