![Cannon TE-BBR Instruction & Operation Manual Download Page 76](http://html1.mh-extra.com/html/cannon/te-bbr/te-bbr_instruction-and-operation-manual_493056076.webp)
CANNON Instrument Company | Troubleshooting
65
TE-BBR Force Reading Error
This section provides information on correcting the Force Reading 0.0 Error relating to the CANNON
Thermoelectric Bending-Beam Rheometer.
The reason for the force reading of 0.0 when the deflection and temperature settings are correct is due
to a failure of the registry of the computer running the instrument. The customer will need reload the
program.
Procedure
1.
Record all information from the certificates for the temperature probe, gauge block, thick and thin
beam.
2.
Remove the Cannon TE-BBR program from the computer being used to run the instrument.
3.
Go to
RUN
type in
REGEDIT
and find the registry keys associated with the TE-BBR. The path is
H_KEY LOCAL MACHINE/SOFTWARE/CANNONINSTRUMENT/BBRw/1.0
4.
Delete the
BBRw
key.
5.
Reload the Cannon TE-BBR program.
6.
Re-enter the data from your certificates.
7.
Recalibrate the instrument.
For a 64 bit Windows Computer
HKEY_LOCAL_MACHINE/SOFTWARE/WOW6432NODE/CANNONINSTRUMENT/BBRW
For a 32 bit Windows Computer
HKEY_LOCAL_MACHINE/SOFTWARE/CANNONINSTRUMENT/BBRW
If the problem is still unresolved, contact CANNON for technical assistance.
Summary of Contents for TE-BBR
Page 1: ...Instruction Operation Manual 44 0991 Thermo Electric Bending Beam Rheometer TE BBR ...
Page 2: ......
Page 10: ......
Page 11: ......
Page 44: ...CANNON Instrument Company Software Menu Options 33 Figure 17 Toolbar Options ...
Page 47: ...36 TE BBR Instruction and Operation Manual Figure 18 Mold Specimen Diagram ...
Page 70: ...CANNON Instrument Company Troubleshooting 59 Figure 25 Diagram A new Figure 26 Diagram B old ...
Page 85: ...74 TE BBR Instruction and Operation Manual Figure 31 Graph View ...