24
TE-BBR Instruction and Operation Manual
5.
Choose a default data path:
a.
Click on
Archive
from the tab options on the
Instrument Setup
property sheet. The current
archive directory is displayed.
b.
Click on
Browse
to select an alternate directory.
c.
Click on the desired directory from the Windows directory dialog.
6.
Change default settings for gage block depth:
a.
Click on
Gage Block
from the tab options on the
Instrument Setup
menu.
b.
Input the gage block height, as measured from the bottom of the block, at each level
(default settings are 16 mm for the top level, 15 mm for the Red level, 13 mm for the White
level and 10 mm for the Blue level).
7.
Change the default settings for the steel beam dimensions:
a.
Click on
Instrument Setup
from the Setup
menu and select
Beam Dimensions
.
b.
Measure the steel beams and input each measured value (up to two decimal places) into the
appropriate field. (These measurements are provided with the beams sent with the
instrument.)
c.
Click
OK
.
8.
Change the default settings for the confidence test tolerance (optional):
a.
Click on
Confidence Test
from the tab options on the
Instrument Setup
menu.
b.
Input the values for minimum and maximum acceptable limits for measurement of the
modulus of elasticity. Default values are 183 GPa for the minimum modulus value, and 223
GPa for the maximum modulus value (203 GPa ± 10 percent).
9.
Add the rest of the default values for the X (horizontal) and Y (vertical) axis of the line graph
created from the BBR test data:
a.
Click on
Graphs
from the tab options on the
Instrument Setup
menu.
b.
Input the values for deflection per graph division (the default is 1 mm) and force per graph
division (the default is 1000 mN).
c.
If desired, click on the Autoscaling option. When checked, the BBRw software will
automatically select graph scale values for effective display of test data.
Summary of Contents for TE-BBR
Page 1: ...Instruction Operation Manual 44 0991 Thermo Electric Bending Beam Rheometer TE BBR ...
Page 2: ......
Page 10: ......
Page 11: ......
Page 44: ...CANNON Instrument Company Software Menu Options 33 Figure 17 Toolbar Options ...
Page 47: ...36 TE BBR Instruction and Operation Manual Figure 18 Mold Specimen Diagram ...
Page 70: ...CANNON Instrument Company Troubleshooting 59 Figure 25 Diagram A new Figure 26 Diagram B old ...
Page 85: ...74 TE BBR Instruction and Operation Manual Figure 31 Graph View ...