Check Function of TE-BBR Cooling Modules
This technical note provides a preliminary procedure for checking function of the thermoelectric cooling
modules mounted internally to the TE-BBR bath. The procedure will check the integrity of the electrical
circuit defined by input and output leads from the modules.
Tools Required:
Phillips screwdriver, multimeter
Procedure
1.
Remove power from the TE-BBR unit and unplug the MAINS power source.
2.
Using a Phillips screwdriver, remove the four screws securing the left rear side panel of the TE- BBR.
Remove the panel to reveal the internal electronics. Refer to Figure 27.
Caution:
The TE-BBR is designed with a capacitor capable of holding a strong
charge. Do not touch the capacitor or other internal components unless
specifically instructed to do so in this procedure.
Figure 27: TE-BBR Capacitor
3.
Locate the internal barrier block. The block includes six pairs of terminal connections. Refer to Figure
28.
Figure 28: Terminal Connections
Summary of Contents for TE-BBR
Page 1: ...Instruction Operation Manual 44 0991 Thermo Electric Bending Beam Rheometer TE BBR ...
Page 2: ......
Page 10: ......
Page 11: ......
Page 44: ...CANNON Instrument Company Software Menu Options 33 Figure 17 Toolbar Options ...
Page 47: ...36 TE BBR Instruction and Operation Manual Figure 18 Mold Specimen Diagram ...
Page 70: ...CANNON Instrument Company Troubleshooting 59 Figure 25 Diagram A new Figure 26 Diagram B old ...
Page 85: ...74 TE BBR Instruction and Operation Manual Figure 31 Graph View ...