CANNON Instrument Company | Unpack & Assemble
11
Figure 4: CT-500 Top Cover with Viscometer Holes
CT-500 Motor Stirrer Installation
Follow the procedure in this section to install the CT-500 Motor Stirrer. If you have the CT-600 Series
instrument, skip to the CT-600 Series installation section and complete the installation directions in
that section manual.
1.
Remove the screws that fasten the top cover to the motor housing and remove the top cover.
2.
Remove the motor-stirrer from its box.
Caution:
Use care when handling the motor stirrer and impeller to prevent
damaging sensitive components. Do not hold the motor assembly by the shaft as
this could bend the rotor shaft out of alignment.
Note:
Two standoffs, one on each side of the opening for the motor-stirrer,
serve as locating pins for the motor support pad. The holes in the pad fit loosely
over these standoffs to prevent the motor from vibrating out of position.
3.
Carefully slide the motor-stirrer into place on the motor support pad, impeller shaft down, and
align the motor line cord so it points toward the rear of the bath. The
motor-stirrer should now lie
flat on the top of the bath.
4.
Lower the motor-stirrer power cord down through the vertical channel in the rear of the unit.
Summary of Contents for CT-500
Page 1: ...25 9991 Instruction Operation Manual Constant Temperature Bath CT 500 600 Series ...
Page 2: ......
Page 4: ......
Page 8: ......