22
CT-500 Instruction Manual
The commands in those areas will be printed in bold uppercase type, with the abbreviation
preceding. For example,
TA - FIXED
refers to the left-hand dial (marked
FIXED
) in the
TEMPERATURE
ADJUST
(
TA
) section of the control panel.
Turning on the CT-500
Temperature Adjust (TA) dials on the CT-500 are equipped with locks to prevent accidental changing
of dial settings. Push the lock up to release the dial. When the dial has been set to the proper
position, push the lock down to re-lock the dial.
1.
Verify that the CT-500 has been assembled and installed correctly per instructions in previous
sections.
2.
Turn the bath Power switch ON. Remember that the bath heaters will not be activated if there is
insufficient liquid in the bath.
Initially the Limit Control (
LC - TEMPERATURE ADJUST
) dial must be turned completely clockwise to
enable the bath to heat properly. After the bath reaches operating temperature, this dial should be
reset to prevent the bath from overheating.
3.
Move the
TA - SELECT
switch to the left (fixed) position.
4.
Release the dial lock on the
TA - FIXED
dial and reset it to
5
.
5.
Re-lock the dial.
Setting the Temperature
There are two ways to set the CT-500 bath temperature. The procedure depends upon whether or
not the desired temperature is on the
T - SELECT
dial.
Warning:
Do not touch hot surfaces on the CT-500 bath when operating the
equipment as injuries and burns may result.
Pre-Set Temperature Selection
1.
Turn the
T - SELECT
dial to the appropriate temperature. If the temperature is above the current
bath temperature, the CT-500 will begin heating and the
HEAT
LED (Light-Emitting Diode) will
light.
2.
Make sure the
TA - SELECT
switch is set to the left (
FIXED
) position.