CT-500 Instruction Manual
Safety Precautions
Always observe these signs and instructions. You must observe cautionary messages and warnings in
order to protect yourself as well as prevent others from physical injury or property damages. The
CT-500 comes equipped with a number of safety features as listed below.
Temperature Fault Sensor
A second thermistor in the temperature bath senses any over-temperature fault condition. If such a
condition occurs, all power is removed from the bath until an operator corrects the problem and
resets the over-temperature limit circuit
Thermistor Detection/Cutoff
If the control thermistor is disconnected, all power to the bath heaters is
cut off.
Bath Fluid Level Cutoff
The CT-500 will not operate if the liquid level in the bath is too low. If such a condition occurs, all
power is removed from the bath until the liquid level in the bath is restored to the minimum safe
level.
Caution:
Safety devices may be impaired if equipment is not operated per
manual instructions.
•
Only qualified personnel should operate the CT-1000.
•
Read and understand all operating instructions and safety precautions listed in this manual
before installing or operating the instrument. Any questions regarding instrument operation or
documentation should be referred to Cannon Instrument Company.
•
Do not deviate from the installation, operation, or maintenance procedures described in this
manual. Improper use of the CT-500 may result in a hazardous situation and may void the
manufacturer’s warranty.
•
Handle and transport the unit with care. Sudden jolts or impacts may cause damage to
components.
•
Never operate the instrument without proper levels of approved bath fluid in bath.
•
Observe all warning labels. Never remove warning labels.
•
Never operate damaged or leaking equipment.