26
CT-500 Instruction Manual
Verifying Limit Control Operation
The Limit Control should be checked periodically to ensure its functionality. To check the Limit
Control, follow the procedure below in this section.
1.
Power on the CT-500 and set the instrument to a desired bath temperature. Wait for the bath to
stabilize at the desired temperature.
2.
Using a screwdriver, slowly turn the Limit Control counter-clockwise until the Limit Control OVER
TEMP warning lights up.
Caution:
If the Limit Control fails to light up and the Limit Control has been
adjusted counter-clockwise to its furthest setting, the unit may need repairs.
Contact CANNON for assistance.
3.
When the Limit Control OVER TEMP warning lights up, turn the Limit Control clockwise for ¼
turn and press the OVER TEMP button to reset the Limit Control for normal operation.
Summary of Contents for CT-500
Page 1: ...25 9991 Instruction Operation Manual Constant Temperature Bath CT 500 600 Series ...
Page 2: ......
Page 4: ......
Page 8: ......