![background image](http://html1.mh-extra.com/html/campbell/skyvue-8m/skyvue-8m_product-manual_489150075.webp)
where
scale (5 characters) = Attenuated_SCALE parameter, %, 0 to 99999. 100% is default
res (2 characters) = Backscatter profile resolution in meters
n (4 characters) = Profile length
energy (3 characters) = Laser pulse energy, %.
lt (3 characters including l/-) = Laser temperature, degrees C
ti (2 characters) = Total tilt angle, degrees
bl (4 characters) = Background light, millivolts at internal ADC input (0 to 2500)
pulse (4 characters) = Pulse quantity x 1000 (0000 to 9999)
rate (2 characters) = Sample rate, MHz, (00 to 99)
sum (3 characters) = Sum of detected and normalized backscatter, 0 to 999. Multiplied by scaling
factor times 10
4
. At scaling factor 100 the SUM range 0 to 999 corresponds to integrated
backscatter 0 to 0. srad
-1
.
CR LF = Carriage Line Feed
LINE 4
SSSSSSSSSSSSSSSSSSSSSSSSSSSSS........(2048 x 5 bytes) CR LF
The two-way attenuated normalized backscatter profile (
CR LF = Carriage Line Feed
LINE 5
ETX CRC-16 EOT CR LF
where
ETX = End-of-Text character
CRC-16 (4 characters) = CRC-16 Checksum
EOT = End-of-Transmission character
CR LF = Carriage Line Feed
SkyVUE™8M Tactical LIDAR Ceilometer
70