Rotate the vial in the sample chamber about 45°.
Place the cover on the instrument.
Press [Read / Avg] or [Read] key.
Note the displayed result.
Repeat this procedure as the smalest NTU value is reading.
Mark the vial.
Indexing more vials repeat point 1. to 12. for each vial.
Repeat this procedure until the reading matches the first vial reading within ± 0,01
NTU.
Mark the vial.
Repeat this procedure to match other cells.
Note:
It may not be possible to match all cells due to variability in glass.