3.6 Trouble-shooting
3.6.1 Operating messages in the display / error display
Display
Meaning
Trouble shooting
Overrange
Reading is exceeding the range
too much light on the photo
cell.
If possible dilute sample.
Cell chamber lid closed?
Repeat measurement
with closed lid.
•
•
•
Underrange
result is under detection limit
– – –
Storagesystem error
use Mode 34
mains power fails or is not
existing.
Insert or change Lithium
battery.
Delete data with Mode 34.
Capacity of
rechargeable battery
Full Capacity
Warning automaticly every 3 min.
Warning automaticly every 12 min.
Warning automaticly, the
instrument switches of
Capacity of rechargeable
battery may be too low.
Charge the rechargeable
batteries.
Operate imstrument with
mains adapter.
•
•
•
Printer „Timeout“
Printer switched off.
No conection.
•
•
Connect printer.
Check connections.
Switch printer on.
•
•
•
Instrument can be
operated with the
mains adapter but
not with the rechar-
geable batteries.
Rechargeable batteries are not
charged or defect.
Fuse (Type A, inert, 20 mm) may be
defect.
Charge rechargeable batteries
or change them.
If the problem still exists chan-
ge fuse.