IB0147 Rev. 08
LEFM 200 Modbus User Manual
Section 3
Page 20
December 2010
Setup Variable Holding
Register
Variable
Definition
Address*
Notes
Kfactor
Float
334
See Installation, Operation, and
Maintenance manual for table of
pulses per unit volume.
When changing to a non-standard k-
factor the desired flow range must
produce an output between these
values:
Minimum Frequency = 3 Hz
Maximum Frequency = 10 kHz
Force a Frequency Output
Integer
802
0 = Normal Operation
Any other value outputs a pulse
output at a frequency equal to the
value entered.
3.3
Display Output Screen Format Definitions
Setup Variable Holding
Register
Variable
Definition
Address*
Notes
FLOW FIELD WIDTH
Integer
776
See Display Notes below
(Section 3.8)
FLOW DECIMAL
PRECISION
Integer
777
See Display Notes below
(Section 3.8)
FLOWUNITS
Integer
784
See Display Notes below
(Section 3.8)