LEFM 200 Modbus User Manual
IB0147 Rev. 8
December 2010
Page 19
Section 3
3.0 LEFM 210E (SINGLE PATH EXTERNAL FLOW METERS)
*Other address locations not listed here were used by Cameron to customize each meter for its
manufactured dimensions.
Note: Integer values are limited to ±32767.
3.1
Registers to Save the Setup
Setup Variable Holding
Register
Variable
Definition
Address*
Notes
Command to use latest entries
BURN AND USE
Integer
170
Activate new setup
values
3.2
Registers for Scaling Flow, Flow Units and Pulse Output
Setup Variable Holding
Register
Variable
Definition
Address*
Notes
MINIMUM FLOW CUTOFF
Float
44
Flow rates below this value ( in
absolute value) are clamped to 0
(display and pulses) and the flow
meter’s totalizers do not update.
UNITS CONVERSION
(converts default cubic feet to
other units)
Float
52
Typically delivered in customer
requested units:
Typical Units:
0.178095238 converts to barrels
2.8317E-02 converts to cubic meters
FLOW RATE TIME UNITS
(converts flow rates from per
second to other time units)
Float
54
Typically delivered in customer
requested units (standard is 3600 for
flow per hour other choices are 1 for
per second 60 for per minute)
Summary of Contents for LEFM 200
Page 4: ...IB0147 Rev 08 LEFM 200 Modbus User Manual Table of Contents Page ii December 2010 ...
Page 22: ...IB0147 Rev 08 LEFM 200 Modbus User Manual Section 2 Page 18 December 2010 ...
Page 28: ...IB0147 Rev 08 LEFM 200 Modbus User Manual Section 3 Page 24 December 2010 ...
Page 40: ...IB0147 Rev 08 LEFM 200 Modbus User Manual Section 4 Page 36 December 2010 ...