CT-BOX Viewer – Quick Start Guide
9
•
concerning the
Oscilloscope
operation mode, the user can set the
sampling
time
(expressed in [
s], N.B. resolution is 10
s).
In the same tab and for both operation modes, by changing the
Ip-turns
parameter, the
user can set the scaling factor for the primary current readings (i.e. the number of
windings of the primary conductor fed in the hole of the DCCT head).
N.B. Once back on the main window by clicking on an empty space (see
),
the highlighted mode at the bottom of the viewer must agree with the one just selected
in the
Acquisition
tab of the
Settings
.
Figure 6:
Acquisition tab settings.
I/O ALARM
In the
I/O ALARM
), the user can set both an
I/O Trigger
and the
limit values to generate an
alarm
, namely:
•
Trigger Source Settings
: when I/O Trigger checkbox is selected, there is the
possibility to choose either the
Generation of an Output Trigger
or the use of
External Trigger
to start the acquisition;
•
Alarm
: when selected, it enables the alarm and the possibility to set both
Upper
and
Lower
limits
(both expressed in [A]) for the read current.