6.9.1.3.2 Calibration using the sensor test and calibration certificate
This method involves using the test and calibration certificate to enter the sensor
data directly in the Sensor Profibus Module. All necessary calibration data can be
found from the certificate.
About the values:
...
These values are adopted directly from the test and calibration
certificate.
This value must be calculated as follows:
Known:
Sensitivity in reference direction
= 1.0538 mV/V
Reference excitation voltage
= 5 V
Required:
Output signal at 100% load = upper calibration value
Sensitivity in reference direction x reference excitation = upper
calibration value
in this case:
1.0538 mV/V x 5 V =
5.269
mV