Test protocol of
an inspection of
electro-medical
equipment
according to
DIN VDE 0751-1:
2001-10
Page 2 of 2
III. Performance check
OK
Not
OK Defect Description
Performance check of the electrial components
Motion limit switching of the motors
Automatic cut-off
Hand switch, blocking functions
Test according to instructions for use
Drive motors
Noise level
Drive motors
Test according to instructions for use
Strain relief and security of routing the plug
Secure, Tight fit
Strain relief for the transformer connecting cable
Secure, Tight fit
Performance check of the mechanical components
Joints and pivots
Smooth operation
Castors
Drive and brake
Emergency lowering of the backrest
Test according to instructions for use
Side rail
Safe locking / unlocking
Lower leg rest
Latching
Accessories (e. g. trapeze pole, trapeze)
Secure, Damage, Suitability
Result of the inspection:
All values within permissible range:
yes
no
Next inspection date:
Inspection passed:
yes
no
If inspection was not passed:
Defective, do not use the bed! => Repair
Defective, do not use the bed! => Take out of Service
Bed does not meet the safety standards
Test approval sticker applied:
yes
no
Part of this inspection report:
Comments:
Inspected on:
Inspector:
Signature:
Evaluated on:
Operator / Expert:
Signature: