38
G
EBRAUCHSANWEISUNG
P
FLEGEBETT
INOVIA · I
NSTRUCTION
M
ANUAL FOR
N
URSING
B
ED
INOVIA
Test protocol of
an inspection of
electro-medical
equipment
according to
DIN VDE 0751-1:
2001-10
Page 1 of 2
Customer / med. Facility / practice:
Address:
Carried out:
Repeat inspection
Inspection prior initial operation
_____________________________________________
Inspection following repairs / servicing
Equipment type:
Hospital bed
Nursing bed
Protection classification:
I
II
Bed type:
INOVIA
Inventory Number:
Location:
Serial Number:
Manufacturer:
Burmeier GmbH & Co. KG
User-specific:
none
Testing equipment used (Type / Inventory Number):
1.
MPG Classification:
Class I
2.
I. Visual Inspection
OK
Not
OK Defect Description
Visual inspection of the electriacl components
Stickers and type plates
Presence, legible
Transformer unit housing
Damage
Motor housing and lifting rods of motors
Correct position, Damage
Hand switch
Damage, Foil intact
Motor-, hand switch- and power cables
Damage, Routing
Visual inspection of the mechanical components
Stickers and type plates
Presence, legible
Trapeze pole / socket
Damage, Deformation
Bed frame
Damage, Deformation
Castors
Damage
Reclining surface
Damage, Deformation
Wooden boards
Damage, Splinters
Welded seams
Split welded seams
Side rail
Damage, Splinters
Knurled screws
Tight fit
Wearing parts such as joints
Damage
II. Electrical measurement of the external transformer unit according to VDE 0751-1:2001-10
(leakage current of equipment: direct or differential)
Proceed as follow:
OK
Not
OK Defect Description
1. Plug in transformer unit in test socket of testing device
2. Insert the plug of the 24 volt cable into the special adapter
(extra accessory; request from BURMEIER)
3. Connect the special adapter with the test probe of the test unit.
4. Start procedures with the measuring instrument
Limit Value
Value of first
measurement
Current actual
value
Result: Bed SK II (Type B)
0,1 mA
mA
mA