50
English
SE58
L
MENU 6 - INPUTS
When one of this function is enabled, the related negative totalizer may be reset through the on/off input.
Totalizers counting lock command enable. When this function is active, applying a voltage on the on/off input
terminals the system stops the totalizers no matter which is the flow rate.
When this function is active (ON), applying a voltage on the on input terminals, the measurement is stopped, the
meter will display zero flow.
When this function is active, applying a voltage on the on/off input terminals the meter performs a autozero
calibration cycle. ATTENTION: If the voltage pulse is less than 1 sec., the meter performs a calibration cycle to
compensate possible thermal drifts. If the voltage pulse is more 1 sec, the meter performs a zero calibration
measure. To perform the calibration it is absolutely necessary for the sensor to be full of liquid and that the
liquid is perfectly still. Even very small movement of the liquid may affect the result of the calibration, and,
consequently, the accuracy of the system.
Range change external command enable. When this function is enabled, applying a voltage on the on/off input
terminals the meter switches to the second measuring range (Fs2).
N.B.: the autorange doesn’t allow to use the manual range change see (POS. 4.4 see page 35).
(POS. 6.3-4) Volume Totalizer Total /Partial Negative reset Enable [T/P-/RESET=ON/OFF]
AL3
[VTTNE] [VTPNE]
(POS. 6.5) Totalizers Count Lock Input Enable
[COUNT LOCK= ON/OFF]
AL3
[TCLIE]
When one of this function is enabled, the related positive totalizer may be reset through the on/off input.
(
POS. 6.1-2) Volume Totalizer Total /Partial Positive reset Enable
[T/P+/RESET=ON/OFF] AL3
[VTTPE] [VTPPE]
(POS. 6.6) MeaSure Lock Input Enable
[Meas.lock=ON/OFF]
AL3
[MSLIE]
(POS. 6.7) CALibration Input Enable
[Calibration=ON/OFF]
AL3
[CALIE]
(POS. 6.8) Scale Range Change Input Enable
[RANGE CHANGE=ON/OFF]
AL3
[SRCIE]
WARNING!
This menu is not enabled when the batching function is active.
Summary of Contents for B0A1C2B1A4A0
Page 1: ...Type SE58 L Transmitter Operating Instructions...
Page 8: ...7 English SE58 L Dimensions 200 114 164 167 114 180 290 195...
Page 9: ...8 English SE58 L Remote version 142 28 295 120 7 30...
Page 11: ...10 English SE58 L DIMENSIONS STAINLESS STEEL VERSION Compact Version 171 156 186 146...
Page 12: ...11 English SE58 L Separate Version 236 40 7 90 241 55...