46
English
SE58
L
to set a defined amount of volume to get one pulse.
To set pulse value set the 2 fields, from left to right: 1)unit of measure, 2) numeric value.
The selection is performed by positioning the cursor in the field that has to be modified.
To change the unit type (metric, imperial) see menu 2 units.
The value of Pls1-2 depends on nominal diameter (max. flow rate) units.
The available units are described in the function POS. 1.4 see page 34.
When the pulse value is defined, also the duration of the pulse shall be set. It’s expressed in milliseconds and the
range is from 0.4 to 1250 ms. Be carefull to set a right value of pulse duration, according to the receiver. It shall
be compatible with the external device which receives the pulses.
ATTENTION:
If the pulse is too long receiver’s coils may burn out or if the pulse is too short, the counter may not
be able to recognize the pulses.
(POS. 3.4-3.6) OutPut 1-2 Pulse Value
[Pls1-2= dm³ x.xxxxx]
AL2
[OP1PV-OP2PV]
(POS. 3.5-3.8) OutPut 1-2 Pulse Time
[Tpls1-2= ms x.xxxxx]
AL2
[OP1PT OP2PT]
(POS. 3.8-3.9) OUtput 1-2 Full scale Frequency
[Frq1-2= x.xxxxx]
AL2
[OU1FF-OU2FF]
Frq1 and Frq2 are activated with POS. 7.1 see page 36 and POS. 7.2 see page 36 enabled and set to the
value freq+/-/+-. Setting duration of the pulse generated on channel 1 and 2.
(POS. 3.2) BaTching QuanTitY
[B.qty=xxxxx]
AL2
[BTQTY]
This function is enabled when output1 is set to “BATCHING”.
It allows to set the unit of measure and the batching quantity.
Summary of Contents for B0A1C2B1A4A0
Page 1: ...Type SE58 L Transmitter Operating Instructions...
Page 8: ...7 English SE58 L Dimensions 200 114 164 167 114 180 290 195...
Page 9: ...8 English SE58 L Remote version 142 28 295 120 7 30...
Page 11: ...10 English SE58 L DIMENSIONS STAINLESS STEEL VERSION Compact Version 171 156 186 146...
Page 12: ...11 English SE58 L Separate Version 236 40 7 90 241 55...