SCS 104
7.10 Drying the Moisture Detector
The moisture detector must be dried if moisture enters.
– Close the gas supply.
– Switch off and unplug the device.
– Loosen the swivel nut for the moisture detector connection line and disconnect the line.
– Unscrew the moisture detector counter-clockwise and remove.
– Dry moisture detector.
– Reinsert the moisture detector and carefully tighten the screw connection.
– Connect the connection line and tighten the swivel nut.
– Restore the power and gas supply.
7.11 Calibration of the moisture detector
– When replacing the moisture detectors, they must be recalibrated.
– Be sure dry gas flows through the cooler.
– Select cooler menu and confirm.
– Select menu item moisture detector.
– The display shows (Reset).
– Confirm the display to calibrate the moisture detectors.
For a detailed overview of menu navigation, refer to chapter "Operation and Control".
7.12 Spare parts and accessories
Please also specify the model and serial number when ordering parts.
Upgrade and expansion parts can be found in our catalog.
Available spare parts:
Item no.
Description
91 100 00 051
Front plate microfuse, 5 x 20 mm, 4 A delayed action
91 100 00 010
Sample gas pump microfuse F1, 5 x 20 mm, 1,6 A delayed action
91 100 00 040
Peristaltic pump microfuse F2, 5 x 20 mm, 0,125 A delayed action
41 11 100
Moisture detector FF-3-N, without cable
7.12.1 Spare parts and accessories
Item no.
Description
41 15 0010
Filter element FE-E1, package of 5 pcs.
90 09 162
O-ring for filter AGF-FE-1
4492 0035 012
Norprene replacement hose with angled connections for peristaltic pump 0.3 L/h
CS9X PM E0111
Extension module gas pumps for SCS 104
CS9X GM E0391
Extension module flow meter DK46 for SCS 104