PCS.base
4 Setup and connection
Check the device for damage prior to installation. This/these could be a damaged housing or add-on components visible from
the outside, such as filter and flow meter. Never use equipment with obvious damage.
CAUTION
Damage/health hazard due to heater leakage
Check the device for obvious leaks before every use and at regular intervals. Any leaks
must be repaired prior to using the device. In addition to gas leaks, leaking fluids in par-
ticular can post an electrical and health hazard.
4.1 Installation site requirements
Be sure the equipment is located on a level, solid surface. Also be sure to comply with the approved ambient temperature.
Do not obstruct the convection of the cooler. There must be adequate room between the vent and the next obstacle (at least 10
cm).
CAUTION
Damage to the device
Protect the equipment against dust, falling objects and external impacts.
Stroke of lightning
Outdoor installation is
forbidden
. As a matter of principle, the operator must regard all
applicable standards according prevention of damage due to lightning, which may oth-
erwise damage the device.
4.2 Connecting a gas probe
The sample gas probe connects to the DN 6 hose connection on the device marked
IN
.
The following devices (e.g. analyser) must be connected to the DN 4 hose connection marked
OUT
with a suitable hose.
4.3 Electrical connections
WARNING
Hazardous electrical voltage
The device must be installed by trained staff only.
CAUTION
Wrong mains voltage
Wrong mains voltage may damage the device.
Regard the correct mains voltage as given on the type plate.
The low heat device socket features an on/off switch which cuts off all poles. This must be set to the zero position prior to con-
necting the electrical.
Connect the included low heat device cable to the device and a suitable voltage source. Please note the correct voltage and fre-
quency. The type plate contains any deviating specifications.
PCS.base with condensate trap may be operated on 110-260 V AC, 50/60 Hz voltage. When using the condensate pump the
voltage is 115 V, 60 Hz or 230 V, 50 Hz.