PCS.base
7.2 Safety instructions
– The device must be operated within its specifications.
– All repairs must be carried out by Bühler authorised personnel only.
– Only perform modifications, servicing or mounting described in this manual.
– Only use original spare parts.
CAUTION
Risk due to defective device
Personal injury or damage to property
a) Switch off the device and disconnect it from the mains.
b) Repair the fault immediately. The device should not be turned on again before elim-
ination of the failure.
DANGER
Toxic, corrosive gas/condensate
Sample gas/condensate may be hazardous to health.
a) If necessary, ensure a safe gas/condensate discharge.
b) Always disconnect the gas supply when performing maintenance or repairs.
c) Protect yourself from toxic/corrosive gasses/condensate when performing mainten-
ance. Wear appropriate protective equipment.
CAUTION
Hot surface
Risk of burns
Housing temperatures may be high during operation.
Allow the unit to cool down before performing maintenance or repairs.
7.3 Replacing the main fuse
– Close gas supply.
– Switch off and unplug the device.
– The main fuse is located on the front panel inside the low heat socket. The square fuse holder above the contacts can be re-
moved by prying with a suitable tool.
– Replace the fuse and push the fuse holder back in.
– Restore the power and gas supply.
13
Bühler Technologies GmbH
BE464003 ◦ 11/2017