3 Transport and storage
The device must always be stored and transported in the original transport bag. Operation without transport bag is prohibited.
Only transport the device in the designated position (upright, level). If this cannot be ensured due to logistics (e.g. transport via
shippers or air transport), it’s important to completely empty the condensate trap (versions without condensate trap), as con-
densate could otherwise flow back into the gas lines. Purge the gas path with preferably dry ambient air to allow acidic condens-
ate to escape. Proceed as described in chapter
Condensate
[> page 10].
DANGER
Toxic, corrosive condensate
a) Protect yourself from toxic, corrosive condensate when performing any type of work.
b) Wear appropriate protective equipment.
c) Please note the national safety rules!
The equipment must be protected from moisture and heat when not in use. It must be stored in a covered, dry and dust-free
room at a temperature between -20 °C and 40 °C.
Outdoor storage is
prohibited
. On principle the operator must meet all applicable standards with respect to preventing damage
due to lightning, which could result in sample gas pump damage.
Storage areas must not contain any equipment generating ozone, e.g. fluorescent lighting, mercury vapour lamps, high voltage
electrical equipment.