BCM-MS, BCM-LS

1 Introduction

These quick guide will help you set up the device. Please note
the safety instructions to avoid damage to the health or prop-
erty damage. Please carefully read the original operating in-
structions including maintenance and troubleshooting in-
structions. You will find these on our website at 

www.buehler-

technologies.com

.

If you have questions, please contact:

Bühler Technologies GmbH
Harkortstraße 29
40880 Ratingen
Germany

Tel.: +49 (0) 21 02 / 49 89-0
Fax: +49 (0) 21 02 / 49 89-20

These operating instructions are a part of the equipment. The
manufacturer reserves the right to change performance-, spe-
cification- or technical data without prior notice. Please keep
these instructions for future reference.

1.1 Intended Use

The BCM-MS200 and BCM-LS200 are used to measure and
document changes in the properties of hydraulic and lubricat-
ing media and simultaneously measure the humidity and
temperature. The respective measured values used to detect
changes in the properties as well as the temperature and hu-
midity are detected continuously, saved and can be read out
at any time using a serial port. If the measured values deviate
from a saved reference, this indicates changes which should
be interpreted and investigated further.

Measured changes in the oil parameters can provide informa-
tion about condition changes, e.g. oil aging, refreshing,
changes or water ingress. Damages can then potentially be
detected early or even avoided. This allows taking suitable
measures to avoid serious machine malfunctions and prolong
maintenance and oil change intervals. Furthermore, meas-
ured parameters and changes in these can be used to derive
and document information related to system maintenance
carried out or the use of the prescribed lubricant type.

The following chapters explain which situations changes in
the condition can be detected. The sensor records the follow-
ing physical oil parameters and their time curve:

• temperature
• relative humidity
• conductivity
• relative fluid permittivity
• Liquid level (BCM-LS200 only)

Since the conductivity and relative permittivity in particularly
greatly depend on the temperature, the sensor is able to con-
vert these parameters to a fixed reference temperature. The
sensor continuously measures at various temperatures for
this conversion and uses this data to determine the temperat-
ure gradients of the parameters.

Determining the temperature gradient requires several tem-
perature cycles when commissioning the sensor. During oper-
ation, the temperature gradient is also continuously adjusted
when oil is changed or ages.

1.2 Scope Of Delivery

• Bühler Condition Monitor BCM
• Product Documentation

2 Safety instructions

The equipment must be installed by a professional familiar
with the safety requirements and risks.

Be sure to observe the safety regulations and generally applic-
able rules of technology relevant for the installation site. Pre-
vent malfunctions and avoid personal injuries and property
damage.

The operator of the system must ensure:

• Safety notices and operating instructions are available

and observed,

• The respective national accident prevention regulations

are observed,

• The permissible data and operational conditions are

maintained,

• Safety guards are used and mandatory maintenance is

performed,

• Legal regulations are observed during disposal,
• compliance with national installation regulations.

3 Transport and storage

Only transport the product inside the original packaging or a
suitable alternative.

The equipment must be protected from moisture and heat
when not in use. It must be stored in a covered, dry, dust-free
room at room temperature.

4 Installation and connection

4.1 Dimensions

BCM-LS200: L = 200 mm, Level measuring range = 115 mm

8 pin A Coding (Male)

8 pin A Coding (Male)

Sealing ring

Sealing ring

Level measuring range

Connection dimensions BCM-LS200

Connection dimensions BCM-MS200

Distance from oil sump

9

Bühler Technologies GmbH

BX150104 ◦ 03/2021

Summary of Contents for BCM-LS

Page 1: ...104 Fluidcontrol Kurzanleitung B hler Condition Monitor deutsch 2 Brief Instructions B hler Condition Monitor english 9 Notice de montage Moniteur de condition B hler fran ais 15 Gu a r pida Monitor d...

Page 2: ...mentiert werden Unter welchen Randbedingungen Zustandsver nderungen zu detektieren sind ist in den folgenden Kapiteln beschrieben Der Sensor erfasst die folgenden physikalischen lkenngr en sowie deren...

Page 3: ...r den lzustand charakteristisches lvolumen zu analysieren sollte der Sensor nicht unmittelbar im l sumpf des Tanks angeordnet werden Idealer Weise erfolgt bei Tankeinbau eine Montage in der N he der...

Page 4: ...rechend der Versorgungsspannung des Sensors angepasst sein Der maximale Lastwiderstand kann mit der Formel 6 1 berechnet werden Alternativ steht die hier auf gef hrte Tabelle zur Verf gung Rmax UVerso...

Page 5: ...ebsbereit D Sensoranschluss bei Datenerfassung ber RS232 Schlie en Sie das Sensorkabel mit dem M12 Stecker an den Sensor an Schlie en Sie den 9 pol D Sub Stecker des Kabels an die entsprechende seriel...

Page 6: ...ionen zum Funktionsempfang Funktionsumfang Szenario Notwendige Informationen zur Anlage Konfigurationsbedarf Grundparameter Tem peratur Feuchte P C P40 C40 Durchschnittstempera tur Lastfaktor seit Inb...

Page 7: ...Sie ob der Sensor korrekt gem Einbauvorschriften ein gebaut ist Das l ist stark versch umt berpr fen Sie ob der Sensor korrekt gem Einbauvorschriften ein gebaut ist Eine Ver sch umung ist beson ders...

Page 8: ...m ssen einprogram miert werden Kontaktieren Sie den B hler Technologies GmbH Service Messbereich falsch eingestellt Der Messbereich ist ls pezifisch und muss ein programmiert werden Kontaktieren Sie...

Page 9: ...records the follow ing physical oil parameters and their time curve temperature relative humidity conductivity relative fluid permittivity Liquid level BCM LS200 only Since the conductivity and relat...

Page 10: ...n Avoid foam in the tank When installing the return or flushing pipe please note the flushing pipe must not run dry in any operating situ ation To avoid thermal factors as much as possible the sensor...

Page 11: ...0 100 U V X R 6250 25 A 6 3 H20 H40 in 0 100 U V X R 6666 67 33 33 A 6 4 AH in ppm 0ppm AHScl U V X R ppm AHScl ppm 16 10 3 A AHScl ppm 4 6 5 P P40 1 5 U V X R 266 67 0 3333 1 A 5 mA Teaching in or s...

Page 12: ...here the sensor is connected to the PC When not using a USB to RS232 adapter this is typically COM 1 When using a USB to RS232 adapter a new virtual COM port will be added Select this If necessary ope...

Page 13: ...inal operating instructions on our website at www buehler technolo gies com 7 Service and Repair For a detailed description of the device and information about troubleshooting and repair please refer...

Page 14: ...n of the sensor and verify the data and power cable are properly connected Please note the prescribed pint assign ment Operating voltage outside the prescribed range Please operate the sensor within a...

Page 15: ...ile suivantes ainsi que leur volution dans le temps Temp rature humidit relative Conductivit permittivit relative du fluide Niveau de remplissage uniquement BCM LS200 tant donn que la conductivit et l...

Page 16: ...rement recouvert d huile dans toutes les situations de fonctionnement de l installation Notez en particulier le volume d oscillation du r servoir ou une ventuelle inclinaison Il convient d viter la f...

Page 17: ...re calcul e l aide de la formule 6 1 Le tableau ci dessous est disponible comme al ternative Rmax UAlimentation V 25 V 200 25 Rmax 200 6 1 Rmax en UAlimentation en V 25 9 50 10 100 12 150 14 200 16 Ta...

Page 18: ...u c ble est connect e l interface s rielle correspondante du convertisseur USB RS232 Connectez le bloc d alimentation et le c ble de d tection Branchez votre bloc d alimentation la tension du sec teur...

Page 19: ...l chargement sur www buehler technologies com Configurations n cessaires la r ception op rationnelle Port e fonctionnement sc nario Informations n cessaires concernant l installation be soins en terme...

Page 20: ...ent expans e V rifiez que le capteur est correctement mont conform ment aux ins tructions de montage On peut s attendre ce qu il y ait moussage en particu lier dans les bo tes de vi tesses lorsque la...

Page 21: ...de mesure est sp cifique l huile et elle doit tre programm e Contac tez le service apr s vente de B hler Technologies GmbH Tab 6 Erreur Aucune communication sensorielle avec le terminal hyperthermiqu...

Page 22: ...e se pueden detectar los cambios de estado El sensor regis tra los siguientes par metros f sicos del aceite as como su evoluci n en el tiempo Temperatura Grado de humedad relativa Conductividad Permit...

Page 23: ...ierto de aceite en todas las opciones de funcionamiento del sis tema Preste atenci n especialmente al volumen oscilante del dep sito y a cualquier posible inclinaci n Debe evitar se la formaci n de es...

Page 24: ...rse con la siguiente f rmula 6 1 Tambi n est disponible la siguiente tabla co mo alternativa Rm x USuministro V 25 V 200 25 Rm x 200 6 1 Rm x en Usuministro en V 25 9 50 10 100 12 150 14 200 16 Tab 1...

Page 25: ...al sensor Enchufe el conector D Sub de 9 pines del cable a la inter faz serial correspondiente del transformador USB RS232 Conecte la fuente de alimentaci n y el cable de sensor Despu s conecte correc...

Page 26: ...n del sensor en el manual de instrucciones completo descarga en www bueh ler technologies com Configuraci n necesaria para la ampliaci n de funciones Variedad de funciones Esce nario Informaci n neces...

Page 27: ...a espuma Compruebe si el sensor es t correctamente instala do de acuerdo a las indi caciones de montaje Es esperable la aparici n de espuma especialmente en engranajes y en posicio nes de instalaci n...

Page 28: ...El rango de medici n es espec fico del aceite y de be programarse P ngase en contacto con el servicio t cnico de B hler Techno logies GmbH Tab 6 Error Sin comunicaci n del sensor con el hiperterminal...

Page 29: ...0 BCM LS200 1 2 BCM 2 3 4 4 1 BCM LS200 L 200 mm Level measuring range 115 mm 8 pin A Coding Male 8 pin A Coding Male Sealing ring Sealing ring Level measuring range Connection dimensions BCM LS200 Co...

Page 30: ...C A 6 2 RH 0 100 U V X R 6250 25 A 6 3 H20 H40 0 100 U V X R 6666 67 33 33 A 6 4 AH ppm 0ppm AHScl U V X R ppm AHScl ppm 16 10 3 A AHScl ppm 4 6 5 P P40 1 5 U V X R 266 67 0 3333 1 A 5 mA 6 6 C C40 p...

Page 31: ...USB M12 D Sub9 USB RS232 D RS232 M12 D Sub9 PC E CMSensorDataViewer exe CMSensorConfig exe CMSensorDataViewer CMSensorConfig COM USB RS232 COM 1 USB RS232 COM Windows COM 5 2 RS232 CAN 5 2 1 RS232 RS...

Page 32: ...C40 P40 C40 P40 C40 P40 C40 5 6 www buehler technologies com 7 www buehler technologies com 7 1 9 V 33 V DC 9600 8 1 N N COM1 RS 232 CMSensorConfig RS232 9 V 33 V DC CAN 9 V 33 V DC CAN CMSensorConfig...

Page 33: ...er technologies com B hler Technologies GmbH Harkortstra e 29 40880 Ratingen Deutschland 49 0 21 02 49 89 0 49 0 21 02 49 89 20 1 1 BCM MS200 BCM LS200 BCM LS200 1 2 B hler BCM 2 3 33 B hler Technolog...

Page 34: ...Coding Male Sealing ring Sealing ring Level measuring range Connection dimensions BCM LS200 Connection dimensions BCM MS200 Distance from oil sump 4 2 BCM MS200 200 375 615 5 BCM MS200 R 16 45 4 5 159...

Page 35: ...6 3 H20 H40 0 100 U V X R 6666 67 33 33 A 6 4 AH ppm 0ppm AHScl U V X R ppm AHScl ppm 16 10 3 A AHScl ppm 4 6 5 P P40 1 5 U V X R 266 67 0 3333 1 A 5 A 6 6 C C40 pS m 100pS m 1000100 pS m U V X R 6 6...

Page 36: ...0 100 3 5 www buehler technologies com 5 1 RS232 USB 1590001001 1590001003 CMSensorDataViewer CMSensorConfig USB USB RS232 1590001002 CMSensorDataViewer CMSensorConfig www buehler technologies com A C...

Page 37: ...Node ID 0x280 0x2E4 dez 740 WTPDO2 TPDO3 ID Node ID 0x380 0x3E4 dez 996 TPDO1 Type 255 WTPDO1Type TPDO2 Type 255 WTPDO2Type TPDO3 Type TPDO2 Type TPDO1 Timer 5000 WTPDO1Timer TPDO2 Timer 5000 WTPDO2T...

Page 38: ...BCM MS BCM LS 5 6 www buehler technologies com 7 www buehler technologies com 7 1 9 33 9600 8 1 N N COM1 RS 232 RS232 CMSensorConfig 9 33 CAN 38 B hler Technologies GmbH BX150104 03 2021...

Page 39: ...BCM MS BCM LS 9 33 CAN CAN RS232 CMSensorConfig B hlerTechnologies GmbH B hlerTechnologies GmbH 6 39 B hler Technologies GmbH BX150104 03 2021...

Reviews: