TC-Double (+)
9.2 Technical Data - Options
Analogue Output Cooler Temperature Technical Data
Signal
4-20 mA or 2-10 V
corresponds to -20 °C to +60 °C cooler temperature
Connection
M12x1 plug, DIN EN 61076-2-101
Digital interface technical data
Signal
Modbus RTU (RS-485)
Connection
M12x1 connector, DIN EN 61076-2-101
Technical Data FF-3-N Moisture Detector
Ambient temperature
3 °C to 50 °C
max. operating pressure with FF-3-N
2 bar
Material
PVDF, PTFE, epoxy resin, stainless steel 1.4571, 1.4576
CPdouble Peristaltic Pump Technical Data
Ambient temperature
0 °C to 55 °C
Flow rate
0.3 L/h (50 Hz) / 0.36 L/h (60 Hz) with standard hose
Vacuum inlet
max. 0.8 bar
Pressure inlet
max. 1 bar
Outlet pressure
1bar
Hose
4 x 1.6 mm
Degree of protection
IP 44
Materials
Hose:
Connections:
Norprene (standard), Marprene, Fluran
PVDF
Technical Data Sample Gas Pump P1
Ambient temperature
0 °C to 50 °C
Operating pressure
max. 1,3 bar abs.
Nominal outlet
280 l/h (at p = 1 bar abs.)
Materials in contact with media vary
by configuration
PTFE, PVDF, 1.4571, 1.4401, Viton, PFA
AGF-PV-30-F2 Filter Technical Data
Ambient temperature
3 °C to 100 °C
max. operating pressure with filter
4 bar
Filter surface
60 cm
2
Filter fineness
2 µm
Dead volume
57 ml
Materials
Filter:
Seal:
Filter element:
PVDF, Duran glass (parts in contact with media)
Viton
sintered PTFE
40 Bühler Technologies GmbH
BE440021 ◦ 09/2021
Summary of Contents for TC-Double Plus