TC-Double (+)
Peristaltic Pump
Display →
PuMP
Switching the peristaltic pump on and off.
Parameter range:
NoP
,
oFF
Factory setting:
NoP
Note:
Status switches, "
PuMP
" flashes.
Heat exchanger material selection
Display →
h.chn
Heat exchanger material selection
Parameter range:
tS
(Steel),
t6
(Glass),
tU
(PVDF)
Factory setting:
tS
(cooler without heat exchanger), or respective material per configuration
Exit main menu
Display →
E
Selecting this will return you to display mode.
5.3.3 Submenu 1
Target temperature (Temperature)
Display → Cooler →
temp
This setting determines the nominal temperature for the cooler temperature.
Parameter range:
2 °C to 20 °C (35.6 °F to 68 °F)
Factory setting:
5 °C (41 °F)
Note:
If the temperature is changed the indicator may blink, until the new operating range has been
reached.
This menu item is hidden if the keylock is enabled.
upper alarm limit (alarm high)
Display → Cooler →
A hi
Here you can set the upper threshold for the visual signal and the alarm relay. The alarm limit is set
based on the cooler temperature setting.
Parameter range:
1 °C to 7 °C (1.8 °F to 12.6 °F)
Factory setting:
3 °C (5.4 °F)
Note:
This menu item is hidden if the keylock is enabled.
19
Bühler Technologies GmbH
BE440021 ◦ 09/2021