TC-Double (+)
5.7 Modbus Register
Description
FC
Address Access Data
type
Default
Min
Max
Selection
Resolution Unit
Block temperature measurement
3
2000
R
Float
-
-
-
-
0.5
°C
Block temperature status
3
2002
R
Uint32
-
-
-
Bit 0 := Error
Bit 1..15 := Reserved
Bit 16:= Sensor not calibrated
Bit 17:= Initialization/measurement invalid
Bit 18 := Stabilisation phase
Bit 19:= Load limit reached
Bit 20:= Measurement outside target range
Bit 21..31 := Not used
-
-
Pre-cooling stage measurement
3
2004
R
Float
-
-
-
-
0.5
°C
Pre-cooling stage status
3
2006
R
Uint32
-
-
-
Bit 0 := Error
Bit 1..15 := Reserved
Bit 16 := Sensor not calibrated
Bit 17 := Initialization / measurement invalid
Bit 18 :=
Bit 19 := Load limit reached
Bit 20 :=
Bit 21..31 := Not used
-
-
Block temperature setpoint
3, 16
5000
R/W
Float
5.0
3.0
20.0
-
0.5
°C
Positive alarm tolerance setpoint
3, 16
5002
R/W
Float
3.0
1.0
7.0
-
1.0
K
Negative alarm tolerance setpoint
3, 16
5004
R/W
Float
-3.0
-3.0
-1.0
-
1.0
K
Moisture detector
signal store error
3, 16
9002
R/W
Uint16
2
-
-
1 := No
2 := Yes
-
-
Moisture alarm
signal store
3, 16
9003
R/W
Uint16
1
-
-
1 := No
2 := Yes
-
-
Sensitivity moisture detector 1
3, 16
9004
R/W
Uint16
1 (with FF)
2 (without FF)
-
-
0 := Sensitivity low
1 := Sensitivity high
2 := Moisture detector inactive
-
-
Sensitivity moisture detector 2
3, 16
9005
R/W
Uint16
1 (with FF)
2 (without FF)
-
-
0 := Sensitivity low
1 := Sensitivity high
2 := Moisture detector inactive
-
-
Display unit selection
3, 16
9006
R/W
Uint16
1
-
-
1 := °C
2 := °F
-
-
Heat exchanger type selection
3, 16
9007
R/W
Uint16
2
-
-
2 := Steel
3 := Glass
4 := PVDF
-
-
25
Bühler Technologies GmbH
BE440021 ◦ 09/2021