TC-Double (+)
4 Installation and connection
4.1 Installation site requirements
The unit is only intended for wall-mounted use in enclosed areas. Adequate protection from the weather must be provided
when used outdoors.
Install the unit leaving enough room below the cooler to discharge the condensate. Leave room above for the gas supply.
Be sure to maintain the approved ambient temperature. Do not obstruct the convection of the cooler. The vents must have
enough room to the next obstacle. The distance must especially be a minimum of 10 cm on the air outlet side.
Ensure adequate ventilation when installing in enclosed housings, e.g. analyser cabinets. If the convection is inadequate, we re-
commend aerating the cabinet or installing a fan to lower the inside temperature.
4.2 Installation
Run the gas supply to the cooler with a downward slope. The gas inputs are marked in red and additionally labelled “IN”. The op-
timal gas connection is detailed in chapter Connecting the heat exchanger.
If a large amount of condensate accumulates, we recommend using a condensate trap with automatic condensate drain before
the cooler.
Glass vessels and automatic condensate drains are available for draining condensate for external mounting below the unit.
When using automatic condensate drains, the sample gas pump must be installed ahead of the cooler (pressure operation) to
ensure proper function of the condensate drain.
If the sample gas pump is located at the cooler outlet (suction operation), we recommend using glass condensate traps or peri-
staltic pumps.
Connecting the condensate drains
Depending on the material, build a connecting line with fittings and tubing or hose between the heat exchanger and condens-
ate drain. For stainless steel the condensate drain can be suspended directly to the connecting tube, for hoses the condensate
drain must be secured separately using a clamp.
The condensate drain can be mounted directly to the heat exchanger.
Condensate lines must always be installed with a slope and a minimum inside diameter of DN 6/8 (1/4”).
4.2.1 Flow adapter connection (optional)
When ordering the cooler with the
option moisture detector without filter
, it will be factory installed inside a flow adapter.
The connection between the heat exchanger outlet and the flow adapter inlet does not have tubing included. The connection
G1/4 or NPT 1/4" (flow adapter marked NPT) for the gas inlet/outlet must be carefully and properly connected using a suitable
screw connection. Here the direction of flow is not relevant.
4.2.2 Connecting the moisture detector (option)
When ordering the cooler with
moisture detector option
, it will be factory installed inside a flow adapter, or for the
filter option
installed and connected in the filter head.
4.2.3 Connecting the peristaltic pump (option)
Coolers ordered with attached peristaltic pump already have it installed and wired. Heat exchangers ordered at the same time
are already installed and connected to the peristaltic pump.
The connection via the condensate outlet screw connection on the pump must be carried out carefully and properly.
NOTICE
Installing peristaltic
pumps
CPsingle / CPdouble limits the maximum permissible
oper-
ating pressure
in the system!
Operating pressure ≤ 1 bar
9