GAS 222.21 AMEX

tion box. Per EN 60079-14 and -15, connection for the alarm
output must be connected energy-limited (U

max

 = 30 V, I

max 

=

100 mA). Please see the enclosed terminal diagram for the
connection.

• Only use cables with a temperature resistance of > 100 °C

to connect to power.

• Make sure the connecting cable has sufficient strain relief

(match cable diameter to the seal on the cubic plug).

• Please note, the heating system briefly has high starting

currents (max. 6 A). Use a suitable fuse (8 A). When con-
necting, please also observe the applicable Ex protection
regulations.

NOTICE

The thermal alarm contact must be connected energy-lim-
ited!

(U

max

 = 30 V, I

max

= 100 mA)

WARNING

High voltage

Damage to the device in case of insulation testing

Do not proceed insulation tests with high voltage to the
device as a whole!

Electric strength test

The necessary tests of all assemblies required to be tested
were carried out at the factory (test voltage 1 kV or 1.5 kV de-
pending on component).

To check the electric strength again yourself, only do so on the
respective individual components.

• Now perform the electric strength test against earth.

4.9.1 Heated pressure vessel (optional)

A heated blowback vessel may optionally be used for blow-
back. Heated via self-regulating PTC heating cartridge to pro-
tect against frost.

The electrical (mains connection 115/230 V AC) connects via
the junction box per the enclosed terminal diagram at the
end.

• Only use cables with a temperature resistance of > 100 °C

to connect to power.

• Make sure the connecting cable has sufficient strain relief

(match cable diameter to the seal on the junction box).

• When connecting, also observe the applicable Ex protec-

tion regulations and general warnings in these operating
instructions.

CAUTION

Equipment damage

Cables damaged

Do not damage the cable during installation. Install a strain
relief for the cable connection. Secure the cable against twist-
ing and loosening. Please note the temperature resistance of
the cables (> 100 °C/212 °F).

DANGER

Adiabatic compression during gas blowback (explosion
hazard)!

Adiabatic compression may cause high gas temperatures and
must be checked by the user.

Gas blowback may result in high gas temperatures due to
adiabatic compression. This can cause flammable gases to ig-
nite spontaneously.

a) Blowback of explosive atmosphere / gases is prohibited.

b) Flammable atmosphere / gases (non-explosive) may only

be blown back with nitrogen (inert gas).

5 Operation and Control

NOTICE

The device must not be operated beyond its specifications.

NOTICE

The weather hood must be closed during operation!

WARNING

Housing or component damage

Never exceed the maximum working pressure and temperat-
ure range of the drive.

DANGER

Explosion hazard due to electrostatic discharge

Equipment may only be used where normal operating condi-
tions do not produce frequent flammable, electrostatic dis-
charge.

5.1 Before Start-Up

Before starting the device, verify:

• The hose and electrical connections and the heating tape

are not damaged and installed correctly.

• No parts of the sample gas probe have been removed.
• The protection and monitoring devices are installed and

functional (e.g. flame arrester).

• The gas inlet and outlet on the gas probe are open.
• Ambient parameters are met.
• Probe parts are resistant to media to be conveyed and in

the surrounding area.

• The performance specifications in the type plate are met.
• The voltage and frequency of the heating tape match the

mains values.

• The temperature has an energy-limited connection (U

max

= 30V, I

max

= 100 mA).

• The electrical connections are tight.
• The monitoring equipment is connected and set as spe-

cified.

• All connection cables are installed without strain.
• Precautions have been taken; earthing.
• The junction box cover is closed and the cable gland is

properly sealed.

12

Bühler Technologies GmbH

BX460050 ◦ 01/2020

Summary of Contents for GAS 222.21 AMEX

Page 1: ...460050 Version 01 2020 Ref 998BX460050 Analysentechnik Kurzanleitung Sonden deutsch 2 Brief Instructions Probes english 8 Notice de montage Sondes fran ais 14 Gu a r pida Sondas espa ol 20 chinese sim...

Page 2: ...en aufgrund m glicher adiaba tischer Kompression hoher R cksp ldruck gegen verschmutz ten Filter nicht zul ssig Das Einhalten dieser Bedingungen liegt im Verantwortungsbereich des Betreibers unter Zuh...

Page 3: ...osta tischen Entladungen kommen kann Reinigen Sie Geh useteile aus Kunststoff und Aufkleber nur mit einem feuchten Tuch Funkenbildung Sch tzen Sie das Betriebsmittel vor externen Schlageinwir kungen F...

Page 4: ...en Bringen Sie elektrische Betriebsmittel die zur Instandhaltung ge ffnet werden m ssen m glichst in einen staubfreien Raum Falls dies nicht m glich ist verhindern Sie das Eindrin gen von Staub in das...

Page 5: ...ann zerbrechen Vorsichtig behandeln nicht fallen lassen Nach Anschluss der Messgasleitung ist diese durch die Schelle abzufangen und zu sichern Bei l ngeren Messgasleitungen sind unter Umst nden weite...

Page 6: ...aturbest n digkeit 100 C besitzen Achten Sie auf ausreichende Zugentlastung des An schlusskabels Kabeldurchmesser dem Dichtungsring der W rfelstecker anpassen Beachten Sie dass beim Heizsystem kurzzei...

Page 7: ...t sind keine Teile der Gasentnahmesonde demontiert sind die Schutz und berwachungsvorrichtungen vorhanden und funktionsf hig sind z B Flammensperre der Gasein und Gasausgang der Gasentnahmesonde nicht...

Page 8: ...he probes is prohibited due to possible adiabatic com pression high blowback pressure against contaminated fil ter The operator is responsible for compliance with these conditions taking into account...

Page 9: ...tions do not frequently produce flammable electro static discharge Always clean plastic housing parts and decals with a damp cloth Sparking Protect the equipment from external blows Flame propagation...

Page 10: ...amage to the system If the process holds a risk of flame propagation install a flame arrestor 4 3 Installing the sampling tube optional The sampling tube if necessary with the fitting extension must b...

Page 11: ...pstream from the pressure vessel On probes used to sample flammable gas nitrogen inert gas must be used for blowback Blowback of explosive gases is prohibited NOTICE The operating pressure of the comp...

Page 12: ...g installation Install a strain relief for the cable connection Secure the cable against twist ing and loosening Please note the temperature resistance of the cables 100 C 212 F DANGER Adiabatic compr...

Page 13: ...tor s manual on the included CD or online at www buehler technologies com for maintenance informa tion 7 Service and Repair Please refer to the original operator s manual on the included CD or online...

Page 14: ...la LIE et la LSE avec les sondes n est pas autoris en raison d une possible compression adiabatique pression de r trolavage lev e contre le filtre encrass Le respect de ces conditions est de la respo...

Page 15: ...h tiques du bo tier ainsi que les au tocollants uniquement avec un linge humide Formation d tincelles Prot gez le moyen de production des chocs externes Retour de flammes En cas de risque de retour de...

Page 16: ...a temp rature maximum de surface de la sonde respectez les normes et la r glementa tion en vigueur Placer si possible l quipement de production lectrique de vant tre ouvert des fins d entretien dans u...

Page 17: ...a le re v tement isolant voire les mors isolants dans la zone de la conduite de gaz de mesure Desserrez pour cela les vis de fixation ATTENTION Risque de rupture Le mat riau isolant peut se briser man...

Page 18: ...ment Le raccor dement de la sortie d alarme doit tre fait par le biais d un cir cuit limit en nergie selon EN 60079 14 et 15 Umax 30 V Imax 100 mA Le raccordement se fait selon le plan d implanta tion...

Page 19: ...ifiez que qu les raccordements des tuyaux lectriques ainsi que le ru ban chauffant ne sont pas endommag s et qu ils sont cor rectement mont s aucun l ment de la sonde de pr l vement de gaz n est d mon...

Page 20: ...LSE con las sondas debido a la posible compresi n adiab tica alta presi n de retrolavado contra el filtro sucio El cumpli miento de estas condiciones es responsabilidad del operador con ayuda de su e...

Page 21: ...hispas Los equipos el ctricos solo deben instalarse en lugares en los que en un uso normal no suelan aparecer cargas electroest ticas inflamables Limpie las partes de pl stico de la carcasa y los adhe...

Page 22: ...perfi cie del aparato tenga en cuenta las normas y regulaciones le gales vigentes Si es posible lleve los equipos el ctricos que deban abrirse pa ra su mantenimiento a un espacio libre de polvo Si no...

Page 23: ...capa aislante y quitar tambi n las mordazas aislantes de la zo na del conducto de gas de muestreo Esto se lleva a cabo soltando los tornillos de fijaci n CUIDADO Peligro de rotura El material aislant...

Page 24: ...onexi n se reali za de acuerdo al esquema de asignaci n adjunto Para conectar el suministro el ctrico utilice nicamente cables con una resistencia t rmica de 100 C Aseg rese de que el cable de conexi...

Page 25: ...o est desmontada Los dispositivos de protecci n y control necesarios est n disponibles y operativos p ej supresor de llamas Las salidas y entradas de gas de la sonda de gas de mues treo no est n bloqu...

Page 26: ...49 0 21 02 49 89 20 1 1 ID 25 1 2 GAS 222 21 AMEX 100105177 462222116633000999999 001 CI 1 Div 2 Gps B C D T3 CSA Certificate 1728394 Install per drawing 46 115 Z01 06 4 115 230 V max 360W 230V at 20...

Page 27: ...GAS 222 21 AMEX 3 20 C 50 C 4 F bis 122 F 4 4 1 4 2 4 3 4 4 O O O 90 4 5 4 6 27 B hler Technologies GmbH BX460050 01 2020...

Page 28: ...NPT 1 4 G1 8 G1 4 NPT 1 4 1 1 2 GAS 222 xx ANSI GAS 222 xx AMEX 4 7 1 G3 8 4 7 2 NPT 1 4 4 7 3 6 mm 1 4 6mm 1 4 4 8 3 bar 44 psi 10 bar 145 psi a b 4 9 100 C 212 F EN 175301 803 DIN43650 115 230V 50...

Page 29: ...ax 30 V Imax 100 mA 1 kV 1 5 kV 4 9 1 PTC 115 230 V AC 100 C 100 C 212 F a b 5 5 1 Umax 30 V Imax 100 mA 6 CD www buehler technologies com 7 CD www buehler technologies com 29 B hler Technologies GmbH...

Page 30: ...nd 49 0 21 02 49 89 0 49 0 21 02 49 89 20 1 1 o 25 1 2 Year 2019 CSA GAS 222 21 AMEX 100105177 462222116633000999999 001 Cl 1 Div 2 Gps B C D T3 CSA Certificate 1728394 Install per drawing 46 115 Z01...

Page 31: ...GAS 222 21 AMEX a b c d a b c a b c 3 20 C 50 C 4 F 122 F 31 B hler Technologies GmbH BX460050 01 2020...

Page 32: ...GAS 222 21 AMEX 4 4 1 4 2 4 3 4 4 90 4 5 4 6 32 B hler Technologies GmbH BX460050 01 2020...

Page 33: ...1 3 2 1 DN65 PN6 DN3 150 2 G3 4 NPT 1 4 G3 8 1 6 1 4 2 NPT 1 4 G1 2 NPT 1 4 G1 8 G1 4 NPT 1 4 1 1 2 GAS 222 xx ANSI GAS 222 xx AMEX 4 7 1 G3 8 4 7 2 NPT 1 4 4 7 3 6 1 4 6 1 4 4 8 3 44 psi 33 B hler Te...

Page 34: ...2 21 AMEX 10 145 psi a b 4 9 100 C 212 F 4 EN 175301 803 ex DIN43650 115 230 50 60 EN 60079 14 15 U 30 I 100 100 C 4 6 8 U 30 I 100 1 1 5 4 9 1 115 230 AC 100 C 34 B hler Technologies GmbH BX460050 01...

Page 35: ...GAS 222 21 AMEX 100 C 212 F a b 5 5 1 U 30 I 100 6 www buehler technologies com 7 www buehler technologies com 35 B hler Technologies GmbH BX460050 01 2020...

Page 36: ...power supply Customer alarm output Probe power supply Heater power supply PAV 01 Alarm output M16 x 1 5 mm 4 5 10 mm M16 x 1 5 M16 x 1 5 2 Connection diagram heated pressure vessel Heater Operating vo...