GAS 222.20 AMEX

4 Installation and connection

4.1 Installation site requirements

Sample gas probes are intended for flange mounting.

• Installation site and installation position are determined

based on requirements specific to the application.

• If necessary, the connection piece should be slightly tilted

toward the centre of the channel.

• The installation site should be protected from the

weather.

• In addition, adequate and safe access for installation and

future maintenance work should be provided. Particularly
follow the uninstalled size of the probe tube!

If the probe is transported to the installation site in pieces, it
will first need to be assembled.

4.2 Installation

DANGER

Danger to life and explosion during installation and
maintenance

The unit must not be worked on (assembly, installation,
maintenance) in explosive atmospheres.

DANGER

Explosion hazard

When used in explosive areas

Flammable gasses and dust could ignite or explode.

Never operate the gas probe outside the specifications. Ex-
tracting gases or gas mixtures which are also explosive in the
absence of air is prohibited.

DANGER

Explosion hazard due to ignition of dust

When using the device in a dusty environment, routinely
clean dust from all components.

The ignition temperature resp. smouldering temperature of
flammable dusts resp. dust layers present must be consider-
ably higher than the maximum surface temperature of the
device (observe applicable standards and statutory regula-
tions).

If possible, take the electrical components which must be
opened for repair to a dust-free room. If unable to do so, pre-
vent dust from entering the housing.

DANGER

Explosion hazard due to flame propagation

Severe injuries and damage to the system

If the process holds a risk of flame propagation, install a
flame arrestor.

4.3 Installing the sampling tube (optional)

The sampling tube, if necessary with the fitting extension,
must be screwed in. The probe is then attached to the mating
flange using the included seal and nuts.

4.4 Installing the outlet filter

NOTICE

The outlet filter and the O-ring for the handle must be inser-
ted prior to first startup.

Operating without outlet filter prohibited!

O-ring handle

Attach an O-ring suitable
for the expected ambient
temperature to the
handle.

Attach the outlet filter to
the handle. Then carefully
insert the handle with fil-
ter in the gas probe and
turn 90° to secure.

Verify the handle is
seated correctly. When
seated correctly it locks
onto the filter housing.

4.5 Insulation

On heated probes completely insulate any exposed flange
areas and, if applicable, the connection piece to absolutely
prevent thermal bridges. The insulating material must meet
the application requirements and be weatherproof.

4.6 Connecting the Gas Line

The sample gas line must be carefully and properly connected
using a suitable fitting.

This table provides an overview of the sample gas probe con-
nections:

Probe

GAS 222

Reservoir

PAV01

Ball valve

pneumatic

drive

Control

valve

3/2-way

solenoid

valve

Connecting
flange 

1)

DN65/PN6/

DN3“-150 

2)

Sample gas
inlet

G3/4

Sample gas
outlet

NPT 1/4

Blowback
connection

G3/8

Test gas
connection 

1)

Tube Ø6 mm

Tube Ø1/4 

2)

Filling port

NPT 1/4

Condensate

G1/2

Bypass

NPT 1/4

Control air

G1/8

G1/4

NPT 1/4

Tab. 1: 

Gas Probe Connections (Varies by Model)

1)

 Varies by version.

2)

 Only GAS 222.xx ANSI and GAS 222.xx AMEX

9

Bühler Technologies GmbH

BX460044 ◦ 01/2020

Summary of Contents for GAS 222.20 AMEX

Page 1: ...460044 Version 01 2020 Ref 998BX460044 Analysentechnik Kurzanleitung Sonden deutsch 2 Brief Instructions Probes english 7 Notice de montage Sondes fran ais 12 Gu a r pida Sondas espa ol 17 chinese sim...

Page 2: ...en aufgrund m glicher adiaba tischer Kompression hoher R cksp ldruck gegen verschmutz ten Filter nicht zul ssig Das Einhalten dieser Bedingungen liegt im Verantwortungsbereich des Betreibers unter Zuh...

Page 3: ...osta tischen Entladungen kommen kann Reinigen Sie Geh useteile aus Kunststoff und Aufkleber nur mit einem feuchten Tuch Funkenbildung Sch tzen Sie das Betriebsmittel vor externen Schlageinwir kungen F...

Page 4: ...ss deutlich ber der maximalen Oberfl chentemperatur des Ger tes liegen zutreffende Normen und gesetzliche Regelungen beachten Bringen Sie elektrische Betriebsmittel die zur Instandhaltung ge ffnet wer...

Page 5: ...r Kalibriergasanschlussleitung optional Zum Anschluss der Kalibriergasleitung wird eine Rohrver schraubung 6 mm bzw 1 4 ben tigt Ist der Kalibriergasanschluss mit einem R ckschlagventil be stellt word...

Page 6: ...nschl sse und Heizband nicht be sch digt und korrekt montiert sind keine Teile der Gasentnahmesonde demontiert sind die Schutz und berwachungsvorrichtungen vorhanden und funktionsf hig sind z B Flamme...

Page 7: ...he probes is prohibited due to possible adiabatic com pression high blowback pressure against contaminated fil ter The operator is responsible for compliance with these conditions taking into account...

Page 8: ...tions do not frequently produce flammable electro static discharge Always clean plastic housing parts and decals with a damp cloth Sparking Protect the equipment from external blows Flame propagation...

Page 9: ...ust from entering the housing DANGER Explosion hazard due to flame propagation Severe injuries and damage to the system If the process holds a risk of flame propagation install a flame arrestor 4 3 In...

Page 10: ...box The plugs are prewired to the junction box For safety reasons this connection must not be modified The electrical connections must only be run to the terminals of the junction box The power suppl...

Page 11: ...limited connection Umax 30V Imax 100 mA The electrical connections are tight The monitoring equipment is connected and set as spe cified All connection cables are installed without strain Precautions...

Page 12: ...la LIE et la LSE avec les sondes n est pas autoris en raison d une possible compression adiabatique pression de r trolavage lev e contre le filtre encrass Le respect de ces conditions est de la respo...

Page 13: ...h tiques du bo tier ainsi que les au tocollants uniquement avec un linge humide Formation d tincelles Prot gez le moyen de production des chocs externes Retour de flammes En cas de risque de retour de...

Page 14: ...des poussi res ou couches de poussi res combustibles pr sentes doit tre nettement au dessus de la temp rature maximum de surface de la sonde respectez les normes et la r glementa tion en vigueur Place...

Page 15: ...t n cessaire dans certaines circonstances de pr voir d autres colliers de fixation sur la voie vers le syst me d analyse Apr s avoir branch toutes les conduites et contr l l tanch it l isolation doit...

Page 16: ...tique ESD possible Les moyens d exploitation ne peuvent tre utilis s que quand un fonctionnement normal n entra nant pas la formation fr quente de d charges lectrostatiques risque d allumage 5 1 Avant...

Page 17: ...LSE con las sondas debido a la posible compresi n adiab tica alta presi n de retrolavado contra el filtro sucio El cumpli miento de estas condiciones es responsabilidad del operador con ayuda de su e...

Page 18: ...hispas Los equipos el ctricos solo deben instalarse en lugares en los que en un uso normal no suelan aparecer cargas electroest ticas inflamables Limpie las partes de pl stico de la carcasa y los adhe...

Page 19: ...ncendido del polvo o de las capas de polvo inflamables debe ser signifi cativamente superior a la temperatura m xima de la superfi cie del aparato tenga en cuenta las normas y regulaciones le gales vi...

Page 20: ...tos de gases de muestreo son muy largos en ciertos casos deber n colocarse m s abrazaderas de seguridad en el trayecto hasta el sistema de an lisis Una vez que todos los conductos est n conectados y s...

Page 21: ...e en lu gar en los que en un uso normal no suelan aparecer cargas electroest ticas inflamables 5 1 Antes de la puesta en funcionamiento Antes de la puesta en funcionamiento del aparato compruebe que L...

Page 22: ...49 0 21 02 49 89 20 1 1 ID 25 1 2 GAS 222 20 AMEX 100105177 462222016633000999999 001 CI 1 Div 2 Gps B C D T3 CSA Certificate 1728394 Install per drawing 46 115 Z01 06 4 115 230 V max 360W 230V at 20...

Page 23: ...s 122 F 4 4 1 4 2 4 3 4 4 O O O 90 4 5 4 6 GAS 222 PAV01 3 2 1 DN65 PN6 DN3 150 2 G3 4 NPT 1 4 G3 8 1 6 mm 1 4 2 NPT 1 4 G1 2 NPT 1 4 G1 8 G1 4 NPT 1 4 1 1 2 GAS 222 xx ANSI GAS 222 xx AMEX 23 B hler...

Page 24: ...2 6 mm 1 4 6mm 1 4 4 7 100 C 212 F EN 175301 803 DIN43650 115 230V 50 60 Hz EN 60079 14 15 Umax 30 V Imax 100 mA 100 C 6 A 8 A Umax 30 V Imax 100 mA 1 kV 1 5 kV 5 5 1 Umax 30 V Imax 100 mA 24 B hler T...

Page 25: ...GAS 222 20 AMEX 6 CD www buehler technologies com 7 CD www buehler technologies com 25 B hler Technologies GmbH BX460044 01 2020...

Page 26: ...nd 49 0 21 02 49 89 0 49 0 21 02 49 89 20 1 1 o 25 1 2 CSA Year 2019 GAS 222 20 AMEX 100105177 462222016633000999999 001 Cl 1 Div 2 Gps B C D T3 CSA Certificate 1728394 Install per drawing 46 115 Z01...

Page 27: ...GAS 222 20 AMEX a b c d a b c a b c 3 20 C 50 C 4 F 122 F 27 B hler Technologies GmbH BX460044 01 2020...

Page 28: ...GAS 222 20 AMEX 4 4 1 4 2 4 3 4 4 90 4 5 28 B hler Technologies GmbH BX460044 01 2020...

Page 29: ...DN3 150 2 G3 4 NPT 1 4 G3 8 1 6 1 4 2 NPT 1 4 G1 2 NPT 1 4 G1 8 G1 4 NPT 1 4 1 1 2 GAS 222 xx ANSI GAS 222 xx AMEX 4 6 1 NPT 1 4 4 6 2 6 1 4 6 1 4 4 7 100 C 212 F 4 EN 175301 803 ex DIN43650 115 230 5...

Page 30: ...GAS 222 20 AMEX EN 60079 14 15 U 30 I 100 100 C 4 6 8 U 30 I 100 1 1 5 5 5 1 U 30 I 100 6 www buehler technologies com 7 www buehler technologies com 30 B hler Technologies GmbH BX460044 01 2020...

Page 31: ...ion Diagram M16 x 1 5 mm 4 5 10 mm Customer power supply Customer alarm output Probe power supply Heater power supply PAV 01 Alarm output M16 x 1 5 mm 4 5 10 mm M16 x 1 5 M16 x 1 5 31 B hler Technolog...

Reviews: