5 Putting into operation
46
R-220 Pro Continuous Operation Manual, Version A
5 12
Heating bath setup for 2×1800 W heating element
The heating element holder is labelled with the heating power data The heating bath of the 2×1800 W
version can be filled with different kinds of liquid heating transfer medium Depending on the maximum
heating bath temperature for the target application, a suitable medium must be used
How to set the maximum setpoint temperature
Limiting the maximal temperature set-point to a value below 180 °C can help to eliminate processing
errors, e g when working with heat-sensitive substances
AERATE
Temp
+
Set Temp
SET
Store setting
X
Keep the "SET" button pressed while switching on
the Rotavapor® to enter the setup menu
X
Set the temperature limit via the turning knob The
set value is displayed as bath temperature value
on the display
X
Press "AERATE" button to store the setting and to
leave the menu