5 Putting into operation
55
R-220 Pro Continuous Operation Manual, Version A
Push down the unlock rocker at the left half of the snap flange
Open the snap flange completely and carefully take away the evaporating flask
Using the flask handler
The handler can increase the level of safety when placed on top of the heating bath at flask exchange
However, the flask handler is not suitable as an installation support! The lift mechanism does not stop
automatically when the evaporating flask has reached its individual installation height!
Using the flask handler installed on the heating bath to lift or lower the evaporating flask
can be dangerous for the operator and can cause device damages
!
WARNING
Serious injuries by flying pieces of broken glass
•
Do not use the bath lift to install or deinstall the evaporating flask
•
Do not operate the lift with the flask handler installed
NOTICE
Risk of instrument damage by wrong mains supply
•
External mains supply must meet the voltage given on the type plate
•
Check for sufficient grounding