8 Troubleshooting
72
R-220 Pro Continuous Operation Manual, Version A
E08 n a
Shortcut at vacuum valve
•
Unplug valve and restart system for
causal research
•
Exchange valve if defective
Call service if problem is persistent
Error Addition error indicator on display Possible cause
Remedy
E11 ‘SHIELD OPEN’ message
appears
Shield not closed at the time of
flask rotation start
•
Close bath shield
•
Press the “Aerate” button to confirm
the error message
Call service if problem is persistent
E12 ‘FLOW’ message appears Cooling medium flow insufficient •
Increase cooling water flow and
press the “Aerate” button to confirm
the error message
•
Adjust sensor
•
Check cable and plug of the flow
sensor
•
Exchange sensor if defective or run
evaporation without flow sensor
Call service if problem is persistent
E13 ‘LEVEL 1’ message appears Filling level reached
•
Empty receiving vessel 1 and press
the “Aerate” button to confirm the
error message
•
Check cable and plug of the flow
sensor
•
Exchange sensor if defective or run
evaporation with user controlled
filling level
Call service if problem is persistent